Tuttnauer ELV - Fully Automatic Vertical Autoclaves
The Tuttnauer vertical, top loading laboratory autoclave is available in chamber sizes from 23 to 160 liters. With higher capacity and better performance, Tuttnauer's ELV Laboratory line of fully automatic vertical steam sterilizers provides safe, economical and effective sterilization. The ELV models have an advanced microprocessor based control panel.
Sterilization Applications
- Liquids such as nutrient media and buffer solutions
- Solid items such as pipettes, tubes and filters
- Glassware and plastic articles
- Biohazard sterilization
The various applications can be carried out using the autoclave's standard programs. Additional options ensure more control over results and optimal, safe and repeatable sterilization processes.
Control and Documentation
- Control Features
- High precision control system for perfect sterilization results
- Password protection allows for secure access control
- Independent temperature and pressure monitoring
- Cycle information recovery in the case of power failure or cycle interruption
- Fail Alert - Indicates cycle failure or interruption
- Door Alert - Indicates the door is unlocked
- RS 232 PC Connection Port for direct software updates and remote maintenance
- Built in Printer
- The laboratory line can be supplied with a built-in printer to document the performed cycle, in addition to a comprehensive LCD display. This feature is optional.
- ADMC Software
- The ADMC software allows complete control and monitoring of up to eight sterilizers. The software retrieves data, creates graphs, tables and print outs, controls the autoclave and shows real-time data.
Safety Features
- Door Safety
- The autoclave lids are designed with a number of independent mechanical and digital safety features.
- A safety device prevents the operator from opening the door when the chamber is pressurized
- Steam is not allowed to enter the chamber when the door is open
- A cycle cannot start if the door is open or not properly locked
- The door cannot unlock until liquid temperature reaches the predetermined end temp
- The door cannot unlock until chamber pressure reaches room pressure
- Temperature Activated Door Lock - The door will not open until the temperature is below a specified safe level
- General Safety Features
- Double Independent Monitoring: The combined electronic and mechanical monitoring guarantees accurate results. The operator has two independent means to monitor temperature and pressure.
- Safety Valves: The chamber is equipped with safety valves - if the pressure exceeds the allowed limit the safety valves will discharge
- Built-in Steam Generator Safety: A water level monitoring system maintains a constant water level and ensures safe operation of the heaters
Quality Features
- The chamber is constructed of long lasting 316Ti grade stainless steel with superior corrosion resistance
- The generator is made from stainless steel
- The autoclave automatically switches to standby mode if no buttons or switches are operated for four hours
- The autoclave is designed for easy servicing allowing maintenance access to all components
- Drain Protection: The exhaust drain is mixed with cold tap water that cools the exhaust?s temperature before reaching the drain
Sterilizer Directives and Standards Compliance
- Tuttnauer's high quality laboratory autoclaves comply with the strictest international directives and standards.
- Pressure Equipment: PED 97/23 EEC, EN 10028-7, ASME Code Sec VIII
- Safety: IEC/UL/EN61010-1, IEC 61010-2-040, EN 61326
- Sterilization: DIN 58951 Series, ISO 17665-1:2006
- Quality System: ISO 9001:2000, ISO 13485:2003
Design:
|
vertical
|
Width (autoclave):
|
495 mm580 mm650 mm880 mm
|
Height (autoclave):
|
645 mm745 mm780 mm925 mm960 mm1010 mm
|
Diameter of vessel:
|
250 mm310 mm380 mm500 mm
|
Height or depth of vessel:
|
400 mm490 mm500 mm690 mm700 mm750 mm
|
Vessel material:
|
stainless steel 1.4571 (AISI 316 Ti)
|
Depth (autoclave):
|
335 mm410 mm415 mm500 mm700 mm
|
Volume (autoclave):
|
23 l40 l55 l62 l85 l110 l160 l
|