Dräger PAC 7000

Item number: DRAEG_37015>

EAN: 4064343257091

Category: Dräger

Manufacturer: Dräger

Price on application
Variations

There are variations of this item. Please select the variation you wish.


Dräger PAC 7000

Warns against hazardous gas concentrations
Every working day places the highest demands on gas monitoring. The device reliably warns against hazardous concentrations of carbon monoxide, hydrogen sulphide (including a Low Concentration option), oxygen, carbon dioxide, sulphur dioxide, chlorine, hydrogen cyanide, ammonia, nitrogen dioxide, nitrogen monoxide, phosphine and organic vapours.

Simple calibration and configuration
The bump test mode and sensitivity calibration can be selected via a password-protected menu. The fresh air calibration can be optimally protected using a password. The device is equipped with an infrared interface and can be connected to a PC via a communication cradle in order to activate the desired functions or download stored data.

Innovative miniature sensors
The innovative Dräger XXS sensors react quickly and produce reliable measurement results. Very short diffusion paths inside the device, as well as the sensors' rapid electrochemical reactions, ensure that any gas-related dangers are displayed as soon as they occur. For maximum safety, the sensor is positioned inside the housing such as to allow gas to reach it from above and from the front. This guarantees flawless measurement, even when the gas entry has been accidentally obstructed.

5-year guarantee option
Dräger offers an optional 5-year warranty for the hydrogen sulphide, oxygen and carbon monoxide monitors.

Robust and long-lasting
Protective rubber ensures that the casing of the Dräger Pac 7000 is particularly shock-proof. The device meets the strict dust and water protection requirements specified in IP 68. The electromagnetic resistance has also been optimised. Dräger has designed the Dräger Pac 7000 for unlimited lifetime. When needed, the sensors can be simply replaced.

Fast bump tests with the Dräger Bump Test Station
To perform an automatic function test with a known gas concentration within a matter of seconds, the device is simply inserted into the Dräger Bump Test Station. The display informs the user if the bump test (function test) was successful. If the test is not performed within a certain time frame, the device can be set-up to deactivate itself.

Triple alarm for extra safety
As soon as the adjustable alarm thresholds are exceeded, or values fall below the configured oxygen concentration, the Dräger Pac 7000 immediately issues an audible, visual and vibrating alarm. In addition, the device is also equipped with adjustable TWA and STEL alarms. The user is also warned when the battery is low or a functional error has occurred.
  • Measures carbon monoxide, hydrogen sulphide, or oxygen
  • No limit to the service life of the device
  • Event logger for up to 60 events
  • Service: Easy to replace batteries and sensors
    Replaceable dust and water filters
  • Warranty of 2 years*
  • Rapid functional tests on site
  • Measures carbon dioxide, sulphur dioxide, chlorine, hydrogen cyanide, ammonia, nitrogen dioxide, nitric oxide, phosphine, or organic vapours
  • Optional 5 year warranty* for the H2S, O2, and CO devices
  • Data logger for storage of gas concentrations and events with date and time (120 hours at 1 record per minute)
  • Configurable TWA and STEL alarms
  • Recording maximum concentrations
  • Automatic function test option
  • Estimated sensor service life of 5 - 8 years

* on the electronics, with the exception of the battery

Variations Measuring type Measuring range Alarm threshold
A1/A2 (standard)
Resolution Response time T90
A H2S 1) 0 - 200 ppm 5 /10 ppm 1 ppm 13 sec.
B H2S 2) 0 - 200 ppm 10 /20 ppm 1 ppm 13 sec.
C H2S 0 - 200 ppm customer request 1 ppm 13 sec.
D H2S LC 0 - 200 ppm customer request 0,1 ppm 13 sec.
E O2 1) 0 - 25 Vol.-% 19 /23 Vol.-% 0,1 Vol.-% 20 sec.
F O2 2) 0 - 25 Vol.-% 19,5 /23,5 Vol.-% 0,1 Vol.-% 20 sec.
G O2 0 - 25 Vol.-% customer request 0,1 Vol.-% 20 sec.
H CO 1) 0 - 1999 ppm 30 /60 ppm 2 ppm 11 sec.
I CO 0 - 1999 ppm customer request 2 ppm 11 sec.
J CO 2) 0 - 1999 ppm 35 /50 ppm 2 ppm 11 sec.
K CO LC 0 - 1999 ppm 30 /60 ppm 1 ppm 15 sec.
L CO2 0 - 5 Vol.-% customer request 0,1 Vol.-% 30 sec. (T50)
M CI2 0 - 20 ppm customer request 0,05 ppm 30 sec.
N HCN 0 - 50 ppm customer request 0,1 ppm 10 sec. (T50)
O HCN PC 0 - 50 ppm customer request 0,5 ppm 10 sec. (T50)
P NH3 0 - 300 ppm customer request 1 ppm 10 sec. (T50)
Q NO2 0 - 50 ppm customer request 0,1 ppm 15 sec.
R NO 0 - 200 ppm customer request 1 ppm 10 sec.
S PH3 0 - 20 ppm customer request 0,01 ppm 10 sec.
T OV 0 - 200 ppm customer request 0,5 ppm 20 sec. (T50)
U OV-A 0 - 200 ppm customer request 1 ppm 40 sec. (T50)
V SO2 0 - 100 ppm customer request 1 ppm 15 sec.
W H2S 3) 0 - 100 ppm customer request 1 ppm 13 sec.
X O2 3) 0 - 25 Vol.-% customer request 1 Vol.-% 20 sec.
Y CO 3) 0 - 1999 ppm customer request 2 ppm 11 sec.



1) Standard configuration europe
2) Standard configuration North America
3) 5-years warranty