GRANT Optima heated, circulating water baths, T100 series

The GRANT Optima heated, circulating water baths of the T100 series are the optimal solution for demanding temperature control in modern laboratory environments. Thanks to their outstanding precision, robustness and flexibility, these laboratory water baths are ideal for applications in life sciences, research, chemistry, industry and materials science. Their advantage over many other water baths available on the market lies in a combination of first-class temperature stability, intuitive operation and versatile configurability, which reliably meets advanced laboratory requirements.

Precise temperature control for professional laboratories

The T100 series from GRANT is characterised by sophisticated technology and durable design. With various models and tank materials - from high-quality plastic to stainless steel - the device can be adapted to specific laboratory requirements. The wide range of volumes from 5 to 38 litres makes the series universally applicable, from the smallest sample incubations to the temperature control of large-volume preparations.

Key features and technical data

  • Temperature range: Depending on the model, from ambient +5 °C / +15 °C or 0 °C (with cooling accessories) to max. 99 °C or 100 °C
  • Temperature stability (at 70 °C according to DIN 12876): ±0.05 °C
  • Uniformity (at 70 °C according to DIN 12876): ±0.1 °C
  • Resolution: 0.1 °C (4-digit LED display)
  • Heating capacity: 1.3 kW (230 V), 1.4 kW (120 V)
  • Tank volume: models from 5, 12, 18, 26 to 38 litres, plastic or stainless steel tank
  • Pump capacity: up to 23 l/min, max. pressure 530 mbar

Other relevant features: The device can be individually recalibrated with a maximum of two calibration points and supports three temperature presets for frequently used setpoints. With a liquid level warning system, adjustable over-temperature shutdown and robust safety features, operation is securely protected even during sensitive reactions or continuous use. The T100 series features a compact housing and tank design, with dimensions optimally adapted to the respective volumes and areas of application.

The stainless steel model (ST series) offers improved chemical resistance to common laboratory chemicals such as buffer solutions, hydrogen peroxide or ethanol, while the plastic models (P series) offer an economical solution for standard applications.

Special features

The series scores with the following special features that make everyday laboratory work easier:

  • Precise electronic temperature control and stable, even water circulation
  • Individually configurable liquid level monitor with float mechanism
  • Uncomplicated operation via an easy-to-read, 4-digit LED display
  • Robust plastic or stainless steel baths for maximum flexibility in the choice of chemicals
  • Various safety mechanisms for uninterrupted and safe temperature control of sensitive substances
  • Option of two-point recalibration for accurate temperature measurement

Certifications and standards

The water baths comply with the requirements of the DIN 12876 standard for temperature control, stability and uniformity of liquid temperature control baths. They therefore meet key criteria for standard-compliant laboratory operation in terms of accuracy, reproducibility and traceability of temperature control.

Typical applications in scientific laboratories

Thanks to their precise temperature control and reliability in long-term use, these devices are particularly suitable for the following laboratory procedures:

  • Incubation of biological samples such as cell cultures, bacterial suspensions and enzyme reactions (e.g. restriction digestion with buffer solutions, PCR reactions)
  • Chemical analyses and syntheses where constant, reproducible temperature control conditions are essential (e.g. control of exothermic reactions, synthesis of polysaccharides and peptides)
  • Materials science testing, including resistance and ageing tests on plastics, elastomers and coatings at defined temperatures
  • Quality control in the food and beverage industry, including pasteurisation, temperature control and testing of fruit juices, dairy products or sample extracts

The models are also popular in the pharmaceutical industry and biochemical research for dissolution, solubilisation or stability testing of active ingredients and reagents such as SDS, Tris-HCl, ethanol or glycerine.

FAQ: Important questions and answers about use

How is the temperature measurement recalibrated?

The two-point recalibration is carried out directly on the controller and can be set for internal temperature probes. This ensures maximum measurement accuracy for changing applications.

Can the water baths be used for external circuits?

Thanks to a powerful, adjustable pump, external circulation operation is possible, e.g. for cooling or temperature control of reaction vessels outside the water bath.

What safety standards are implemented?

All models have a liquid level sensor (float switch) and a fixed overtemperature protection system to ensure safe operation at all times and prevent damage.

Are the water baths suitable for media containing solvents?

The stainless steel tank models can be used with many common solutions such as buffers, water, ethanol or other moderately aggressive chemicals (no concentrated acids/bases).

How flexible is the choice of language and display?

The T100 series models use a 4-digit LED display with simple parameterisation - various language options are not available on the device for international use, but the user interface is intuitively designed.

Which programmes can be used to control the temperature?

Three temperature setpoints can be stored, and the desired target temperature can be selected directly depending on the application.

Conclusion

The GRANT Optima T100 series circulating water baths combine first-class precision, excellent long-term stability and a comprehensive protection mechanism, as required by professional laboratories today. Thanks to their flexible configuration, wide range of models and maximum reliability, they are establishing themselves as the standard solution for all temperature control tasks in scientific and industrial environments.


Heating bath model: water/fluid bath
Cooling: no
Model Temperature range
T100-P5 ambient +15 to 99°C
T100-P12 ambient +5 to 99°C
T100-P18 ambient +5 to 99°C
T100-ST5 ambient +15 to 100°C
T100-ST12 0 to 100°C
T100-ST18 0 to 100°C
T100-ST26 0 to 100°C
T100-ST38 0 to 100°C


Heating circulator T100
Stability (DIN 12876) at 70°C (°C) ±0.05
Uniformity (DIN 12876) at 70°C (°C) ±0.1
Setting resolution (°C) 0.1
Display 4 digit LED
Timer function -
No. of temperature presets 3
Re-calibration points 2
Socket for external probe (TXPEP, TXSEP) -
Communications interface -
Programmable -
Relays -
Safety
over temperature
fluid level - float

fixed
+
Language capability -
Alarms (can be configured to switch a relay) -
Heater power
230 V (kW)
120 V (kW)

1.3
1.4
Electrical power
230 V (kW)
120 V (kW)

1.4 (50-60 Hz)
1.5 (50-60 Hz)
Height above tank rim (mm) 200
Depth below tank rim (mm) 135

Tank P5 P12 P18
Material plastic plastic plastic
Capacity (L) 5 12 18
Outer tank dimensions (h x l x w) (mm) 180x240x330 180x415x350 180x600x365
Working area (l x w) (mm) 120x150 210x280 280x325
Min/max liquid depths (mm) 85/140 85/140 85/140
Inner tank dimensions (l x w x h) (mm) 240x160xx150 325x208x150 510x290x150
Overall dimensions incl. controller (l x w x h) (mm) 390x200x380 415x350x380 600x350x380

Tank ST5 ST12 ST18 ST26 ST38
Material stainless steel stainless steel stainless steel stainless steel stainless steel
Capacity (L) 5 12 18 26 38
Outer tank dimensions (h x l x w) (mm) 200x330x180 200x360x330 200x540x330 255x540x330 255x730x330
Working area (l x w) (mm) 150x150 205x300 385x300 385x300 575x300
Min/max liquid depths (mm) 85/140 85/140 75/130** 125/180** 125/180**
Inner tank dimensions (l x w x h) (mm) 300x150x150 325x300x150 505x300x150 505x300x200 690x300x200
Overall dimensions incl. controller (l x w x h) (mm) 330x180x395 360x330x395 540x330x395 540x330x405 730x333x405


**...maximum depth can be increased by 10 mm, by removing the circulation tray in 18, 26, 38 litre baths, with slight loss of performance

Operation at or below ambient temperature requires accessory cooling

Information on product safety
Manufacturer information:
BioSan SIA
Ratsupites 7, build.2
1067 Riga, Latvia, Lettland
[email protected]
Responsible person:
Mr. Schuldt
0049 30 403 667 940