BINDER Constant climate chamber KBF-S 1020Constant climate chambers with large temperature / humidity range
The new Solid.Line KBF-S climate chamber is a custom-made solution for long-term testing and accelerated stability testing for pharmaceutical products in accordance with the ICH Q1A guideline.
The corrosion-free and condensation-free interior also ensures exceptionally stable temperature and moisture levels, even when fully loaded.
Important characteristics
- Temperature range: 0 °C to 70 °C
- Humidity range: 20% to 80% RH
- APT.line preheating chamber technology
- Humidity regulation with capacitive humidity sensor and steam humidification
- LCD to display temperature and humidity along with additional information and alarms
- Internal data logger, measured values can be read out in open format via USB
- Unit self-test for comprehensive status analysis
- Inner chamber made completely of stainless steel
- 2 stainless steel racks, inclusive shelf supports
- Access port with silicone plug, 30 mm, left
- Robust twin casters with brakes, from 240 L
- Class 3.1 independent temperature safety device (DIN 12880) with visual and audible temperature alarm
- Computer interface: Ethernet
- Independent water supply via tank
Load max.:
|
101 - 150 kg
|
Width outside (incubators,...):
|
1250 mm
|
Width inside (incubators,...):
|
976 mm
|
Maximum temperature (heat):
|
+ 70 °C
|
Height outside (incubators,...):
|
1925 mm
|
Height inside (incubators,...):
|
1250 mm
|
Interior volume:
|
1001 - 1500 l
|
Cabinet type:
|
constant climate chamber
|
Depth outside (incubators,...):
|
1145 mm
|
Depth inside (incubators,...):
|
836 mm
|
Lowest temperature (cooling):
|
0 °C
|
Product weight: |
379,00 Kg |
Dimensions: |
× × cm |
BINDER Constant climate chamber |
KBF-S 1020-230V1 |
KBF-S 1020UL-240V1 |
Option model |
Standard |
Standard |
Performance Data Temperature |
|
|
Max. heat compensation at 40 °C [W] |
550 |
550 |
Temperature fluctuation at 40 °C [+ K] |
0.1 |
0.1 |
Temperature range [°C] |
0 to 70 |
0 to 70 |
Temperature variation at 40 °C [+ K] |
0.3 |
0.3 |
Performance Data Climate |
|
|
Humidity fluctuation at 25 °C and 60 % RH |
1.5 % RH |
1.5 % RH |
Humidity range [% RH] |
20 to 80 |
20 to 80 |
Humidty recovery time after 30 s door open at 25 °C and 60 % RH [min] |
11 |
11 |
Humidty recovery time after 30 s door open at 40 °C and 75 % RH [min] |
12 |
12 |
Temperature fluctuation at 25 °C and 60 % RH [+ K] |
0.1 |
0.1 |
Temperature fluctuation at 40 °C and 75 % RH [+ K] |
0.1 |
0.1 |
Temperature range [°C] |
10 to 70 |
10 to 70 |
Temperature variation at 25 °C and 60 % RH [+ K] |
0.3 |
0.3 |
Temperature variation at 40 °C and 75 % RH [+ K] |
0.3 |
0.3 |
Electrical data |
|
|
Nominal power [kW] |
2.0 |
2.0 |
Phase (Nominal voltage) |
1~ |
1~ |
Power frequency [Hz] |
50/60 |
50/60 |
Rated Voltage [V] |
200 to 230 |
200 to 240 |
Unit fuse [A] |
16 |
16 |
Doors |
|
|
Unit doors |
2 |
2 |
Internal Dimensions |
|
|
Depth [mm] |
836 |
836 |
Height [mm] |
1250 |
1250 |
Width [mm] |
976 |
976 |
Measures |
|
|
Interior volume [L] |
1020 |
1020 |
Load per rack [kg] |
45 |
45 |
Net weight of the unit (empty) [kg] |
379 |
379 |
permitted load [kg] |
150 |
150 |
Wall clearance back [mm] |
100 |
100 |
Outer dimensions |
|
|
Depth net [mm] |
1145 |
1145 |
Height net [mm] |
1925 |
1925 |
Width net [mm] |
1250 |
1250 |
Environment-specific data |
|
|
Energy consumption at 40 °C and 75 % RH [Wh/h] |
550 |
550 |
Sound-pressure level [dB(A)] |
56 |
56 |
Fixtures |
|
|
Number of shelves (std./max.) |
2/15 |
2/15 |
1 All technical data is specified for unloaded units with standard equipment at an ambient temperature of +22 °C
+3 °C and a power supply voltage fluctuation of
+10 %.
The temperature data is determined in accordance to BINDER factory standard following DIN 12880, observing the recommended wall clearances of 10 % of the height, width, and depth of the inner chamber.
Technical data refers to 100 % fan speed.
All indications are average values, typical for units produced in series.
We reserve the right to change technical specifications at any time.