The Spetec CleanBoy Mini cleanroom station is a sophisticated, compact laminar flow station for responsible applications in the field of cleanroom technology. Its main task is to efficiently protect sensitive products and processes from airborne particles, making it ideal for applications such as semiconductor manufacturing, optics and laser production, microelectronics and food processing. Compared to other solutions, this station impresses with its particularly space-saving design, modular construction and excellent energy efficiency combined with low noise levels. It thus offers versatile and economical protection at the highest level, especially for users who expect maximum flexibility and uncompromising environmental cleanliness. Innovative features and technical highlights The CleanBoy Mini addresses the key requirements of a modern cleanroom station: uncompromising hygiene, flexible installation and reliable personalisation protection. Optimised for professional applications, this laminar flow station is equally suitable for routine work and highly sensitive special processes. The cleanroom station integrates numerous features to ensure maximum ease of use and operational safety.
Important technical features at a glance:
- Compact device design: choice of five model sizes for optimum space adaptation
- Powerful HEPA filters: filter dimensions from 610 x 610 mm to 1830 x 610 mm, separation efficiency ?99.995% (H14, EN 1822)
- Flexible external dimensions: From 735 x 740 x 1204 mm (Mini 37) to 1955 x 740 x 1204 mm (Mini 112) - identical working height for ergonomic handling
- Low-noise operation: Low noise emission, specially designed for laboratory environments and production cells
- Energy-saving fan control: Reliable air flow with reduced power consumption
- Modular design: Quick filter replacement and easy maintenance thanks to service-friendly design
- Optional accessories: Expandable with UV light, additional lighting or customer-specific adaptations
Thanks to its flexible dimensions and options, the CleanBoy Mini cleanroom station offers enormous versatility for a wide range of processes. Its robust safety glass panes support activities that require maximum product protection and high visibility.
The large filter area ensures consistent cleanroom conditions in accordance with ISO 5 (EN ISO 14644) and GMP, even for demanding applications. The modular design makes maintenance, filter replacement and any subsequent retrofitting particularly efficient - a key argument for anyone who needs to remain flexible during ongoing operations.
Certifications and standards
The CleanBoy Mini is certified in accordance with EN ISO 14644 (cleanroom classes), GMP guidelines, DIN EN ISO 14644 and VDI 2083. It is also TÜV-tested. As part of the service processes, all test parameters are documented and certified in accordance with current standards and ISO specifications, including:
- Particle count (GMP, EN ISO 14644)
- Flow velocity
- Air exchange rate and filter performance
This comprehensive documentation guarantees absolute legal certainty and traceability for every step of the manufacturing process. Specific applications Special application examples highlight the versatility of the CleanBoy Mini:
- Particle-free assembly and quality control of laser optics and lens systems in clean room environments
- Contamination-free handling area for the precision manufacturing of semiconductor wafers (e.g. in connection with lithographic substances, photoresists, chemical etchants)
- Protection of sensitive electronic components during circuit board and microchip production, especially in contact with fine dust or in ESD-prone environments
- Hygiene area for sensitive process steps in food processing, e.g. when packaging open products, when handling additives such as enzyme preparations or starter cultures
The station is also suitable for applications in medical diagnostics, pharmaceutical laboratories, materials engineering and for repair work on precision instruments when it is necessary to ensure that the environment remains free of disruptive particles, germs and aerosols.
Answers to frequently asked questions (FAQs)
Which filters are installed and what class is achieved?
- Built-in HEPA H14 filters in accordance with EN 1822 offer a filtration performance of ?99.995% and create clean room conditions of class ISO 5 (according to EN ISO 14644).
How often should filters be serviced or replaced?
- Maintenance is carried out depending on particle load and as required. An inspection is recommended at least once a year. Filter replacement is easy to carry out and is typically required every 6 to 12 months.
Is continuous operation possible, even in sensitive areas?
- Operation in 24/7 mode is provided for, and energy-optimised fan control minimises costs and emissions. The perfect solution for continuous process protection as well.
What safety features are integrated?
- The robust construction is equipped with single-pane safety glass (ESG), comprehensive electrical safety devices and an emergency stop function. Ventilation testing is carried out during every installation.
What services does the manufacturer offer?
- In addition to consulting and customised selection, there is a certified maintenance and validation service in accordance with EN ISO 14644 and GMP, as well as particle counting, filter replacement, leak detection and volume flow measurement.
Summary
For a wide range of industrial and scientific applications where the highest cleanroom standards and product safety are mandatory, the CleanBoy Mini from Spetec proves to be a future-proof investment. The combination of sophisticated filter technology, a wide range of sizes, modular equipment and reliable certifications in accordance with international standards covers a broad spectrum of professional requirements - with simple maintenance and permanently low operating costs.
|
Width:
|
701 mm - 800 mm1001 mm - 1100 mm1301 mm - 1400 mm1601 mm - 1700 mm1901 mm - 2000 mm
|
|
Device type:
|
Product protectionclean benchfilter module
|
|
Height:
|
1201 mm - 1300 mm
|
|
Cleanroom equipment:
|
Clean room station
|
|
Depth:
|
701 mm - 800 mm
|
| Clean room station |
Spetec CleanBoy Mini |
| Type/Size |
Mini 37 |
Mini 56 |
Mini 75 |
Mini 93 |
Mini 112 |
| Filter dimensions (mm) |
610 x 610 |
915 x 610 |
1220 x 610 |
1525 x 610 |
1830 x 610 |
Outer dimensions B x T x H (mm) |
735 x 740 x 1204
|
1040 x 740 x 1204
|
1345 x 740 x 1204
|
1650 x 740 x 1204
|
1955 x 740 x 1204
|
Spetec Clean room technology Service
We - as a manufacturer and service provider for clean room technology - provide a range of device variants on the one hand, and fast and universal service assistance on the other hand. Now we have used our many years of experience as a manufacturer to develop a services offering, which allows us to test, service, and maintain your clean room system using state of the art measuring equipment.
Very clean air is not just vital in clinic operating rooms. Highly sensitive production environments, such as those in the semiconductor production, precision engineering, optics and optoelectronics, laser technology, or food processing industries, also demand a sufficient level of ambient air quality. In order to ensure this, flow velocity or filter tests are required at regular intervals.
We offer the following services:
- Particle counts
- Replacing filters in various brands of laminar flow boxes (LF units)
- Finding and fixing leaks
- Determining volume flow rates or flow velocities
- Temperature and relative humidity measurements
-HEPA filter / filter system installation monitoring
- DEHS leak test
- Clean room validation and certification in accordance with EN ISO 14644 and GMP guidelines
We also perform certification within the framework of EN ISO 14644, that is, we record the measurement parameters defined in DIN EN ISO 14644 such as particle counts, flow velocities, etc. All of these parameters are confirmed with a corresponding certificate.
All of the particle measurement devices, air speed, temperature, and humidity sensors we use are tested and calibrated every 12 months.
Our offering includes service and maintenance work as well as certification of clean rooms and laminar flow boxes according to the current national and international guidelines and standards. These include the VDI 2083 and DIN EN ISO 14644 guidelines as well as the US Fed Std. 209E. You receive comprehensive documentation about the measurements and a risk assessment.
Info Reinraumstation CleanBoy
Info Clean room station CleanBoy