Spetec Laminar flow module FMS series ''SuSi''

Item number: SPETEC_23040

EAN: 4064343045285

Category: Spetec

Manufacturer: Spetec

starting from 2.277,78 €

Excluding 19% VAT. , plus shipping (Sperrgut)

(including 19% VAT. 2.710,56 €)
module

There are variations of this item. Please select the variation you wish.


Spetec Laminar flow module FMS series ''SuSi''
Filter module without box

The laminar flow module which comes equipped with a ceiling suspension can for example be used as a clean air shower or be integrated directed in a machine enclosure. Through the master/slave function, one control unit is able to control up 100 slave-modules (optional).

Filter change
The main filter is stored in a filter cartridge which is attached to the laminar flow module on its underside. This mechanism is especially advantageous when the modules are suspended from the ceiling or when these are mounted on clean room cells. As a result the filter can easily be changed from underneath.

Application examples
The laminar flow modules are also used on top of clean room cells.

Control panel
The flow velocity can be controlled in 8 steps. The controls include the ''night reduction'' (eco) mode to save energy costs, and the ''MAX'' function to clean the workstation, parts, or equipment if applicable.
  • 8 level adjustment of the flow velocity with night reduction and flushing function
  • LCD display with selected flow velocity in m/second
  • Automatic flow adjustment with filter replacement indicator (optional)
  • LED fault indicator
  • Connection options for building control systems (remote monitoring) (optional)
  • Interior light switch on the control panel (optional)
  • Service and interval indicator, dependent on the usage period
  • Master/Slave function: Position of Control Panel for Laminar Flow Module freely selectable
Through the use of latest generation EC fans with a very high efficiency, the power draw is reduced to a minimum.

Width: 601 mm - 700 mm901 mm - 1000 mm1201 mm - 1300 mm1501 mm - 1600 mm1801 mm - 1900 mm
Device type: Product protectionfilter moduleclean bench
Height: up to 500 mm
Cleanroom equipment: Laminar flow box
Depth: 601 mm - 700 mm
Laminar flow module Spetec FMS series ''SuSi''
Type/Size FMS 24 FMS 37 FMS 56 FMS 75 FMS 93 FMS 112
Filter area (mm) 610 x 400 610 x 610 915 x 610 1220 x 610 1525 x 610 1830 x 610
Weight (kg) 20 28 31 49 56 63
Outer dimensions
W x D x H (mm)
676 x 669 x 430.5 981 x 669 x 430.5 1286 x 669 x 430.5 1591 x 669 x 430.5 1896 x 669 x 430.5
Spetec Clean room technology Service

We - as a manufacturer and service provider for clean room technology - provide a range of device variants on the one hand, and fast and universal service assistance on the other hand. Now we have used our many years of experience as a manufacturer to develop a services offering, which allows us to test, service, and maintain your clean room system using state of the art measuring equipment.

Very clean air is not just vital in clinic operating rooms. Highly sensitive production environments, such as those in the semiconductor production, precision engineering, optics and optoelectronics, laser technology, or food processing industries, also demand a sufficient level of ambient air quality. In order to ensure this, flow velocity or filter tests are required at regular intervals.

We offer the following services:
- Particle counts
- Replacing filters in various brands of laminar flow boxes (LF units)
- Finding and fixing leaks
- Determining volume flow rates or flow velocities
- Temperature and relative humidity measurements
-HEPA filter / filter system installation monitoring
- DEHS leak test
- Clean room validation and certification in accordance with EN ISO 14644 and GMP guidelines

We also perform certification within the framework of EN ISO 14644, that is, we record the measurement parameters defined in DIN EN ISO 14644 such as particle counts, flow velocities, etc. All of these parameters are confirmed with a corresponding certificate.

All of the particle measurement devices, air speed, temperature, and humidity sensors we use are tested and calibrated every 12 months.

Our offering includes service and maintenance work as well as certification of clean rooms and laminar flow boxes according to the current national and international guidelines and standards. These include the VDI 2083 and DIN EN ISO 14644 guidelines as well as the US Fed Std. 209E. You receive comprehensive documentation about the measurements and a risk assessment.

Katalog Reinraumtechnik

Catalog Clean room technology