em-tec BioProTT Clamp-On Transducer

Item number: EMTEC_37004

EAN: 4064343195928

Category: em-tec

Manufacturer: em-tec

Price on application
Transducer-type

There are variations of this item. Please select the variation you wish.


em-tec BioProTTTM Clamp-On Transducer

  • Non-invasive Clamp-On Transducers for flexible tubing
  • Easy to install - no need to splice lines or stop process
  • No pressure drop in system
  • Does not induce additional shear stress
  • Economical - reusable transducers last for years

Modell: BioProTT Clamp-On
Measurement type: Ultrasonic flowmeter
Shipping weight:1,00 Kg
Type Maximum Flow Range 1)Tube size in inchTube size in mm
[l/min]ID x WT ["]OD ["]ID x WT [mm]OD [mm]
BCT 1/8 x 1/16" Aup to ± 41/8 x 1/161/43.18 x 1.596.35
BCT 11/64 x 3/64" Aup to ± 611/64 x 3/6417/644.30 x 1.256.80
BCT 3/16 x 1/16" Aup to ± 63/16 x 1/165/164.76 x 1.597.94
BCT 1/4 x 1/16" Aup to ± 81/4 x 1/163/86.35 x 1.599.53
BCT 1/4 x 3/32" Aup to ± 81/4 x 3/327/166.35 x 2.3811.11
BCT 3/8 x 1/16" Aup to ± 103/8 x 1/161/29.53 x 1.5912.70
BCT 3/8 x 3/32" Aup to ± 103/8 x 3/329/169.53 x 2.3814.29
BCT 3/8 x 1/8" Aup to ± 103/8 x 1/85/89.53 x 3.1815.88
BCT 1/2 x 3/32" Aup to ± 201/2 x 3/3211/1612.70 x 2.3817.46
BCT 1/2 x 1/8" Aup to ± 201/2 x 1/83/412.70 x 3.1819.05
BCT 1/2 x 3/16up to ± 201/2 x 3/167/812.70 x 4.7622.23
BCT 3/4 x 1/8up to ± 203/4 x 1/8119.05 x 3.1825.40
BCT 3/4 x 3/16up to ± 303/4 x 3/169/819.05 x 4.7628.56


1) depending on medium and tube size

  • ID = inner diameter
  • OD = outer diameter
  • WT = wall thickness
  • BCT = BioProTTTM Clamp - On Transducer


Technical Specifications
Size (HxWxL)25 x 33 x 45 mm (tube inner diameter / ID < 1/2")
27 x 38 x 51 mm (tube inner diameter / ID = 1/2")
Total weight (incl. cable/plug)137 g (tube ID < 1/2")
153 g (tube ID = 1/2")
Weight sensor head52 g (tube ID < 1/2")
72 g (tube ID = 1/2")
Housing / lid materialEpoxy resin, aluminium, stainless steel
Cable length2.9 m
IP - CodeIP65
Fluid typesWater, Buffer Solutions, Cell Culture Media, Fermentation Media, Non-Aerated Liquids
Tubing typese.g. Silicon, PVC and Polyurethanes, including Tygon®, C - flex®, Pharmed®, others on request
Medium operating temperature4 to 50° C (40 to 122° F)
Calibration tablesUp to 7 calibration tables can be stored in one sensor
Cleaning and disinfectionDisinfect easily using alcohol based surface cleaners
Compatible flow metersBioProTTTM Flow Track, Flow Track plus, Flow Track DINrail
Accuracy in combination with BioProTTTM flow meters
Accuracy± 3% of reading ± max. 20 ml/min
Reference conditions for above accuracy statement:
Liquid temperatureCalibration temperature ± 5 ° C (± 9 ° F )
Ambient temperature20° C ± 5 ° C (68 ° ± 9 ° F )
Straight inlet section10 x ID of tubing, fully developed flow profile
TubingRaumedic - ECC - Blood Line PVC with acoustic couplant applied