- Volume, Channel
-
There are variations of this item. Please select the variation you wish.
Socorex Acura® electro 956 multichannel pipettes, micropipettes and macropipettes: The microprocessor-controlled instrument includes intuitive software avoiding any complicated programming steps. No particular skills are required to get quick access to six working modes and many volumes to select from. Volumes range from 0.1 µL to 10 mL.
- Optimal ergonomics, light weight
- Easy, intuitive, self-teaching programme
- Large display, reversible right / left reading
- Swift-interchange battery pack
- Fast charge, long working autonomy
- Adjustable tip ejector (Socorex patented) fitting most tips
- Pipetting cycle counter
- Interchangeable volumetric modules
Socorex Acura® electro 956 multichannel micropipettes include a microprocessor-controlled motor and rechargeable battery pack. Eight models with 8 and 12 channel cover the volume range 0.5 to 350 µL. Additional interchangeable volumetric modules available as accessories.
- Large selection of 8- and 12-channel models
- Light weight and perfect hand fitting
- Sequential tip ejection
- Also fit any single channel volumetric module
Acura® electro initial package includes electronic pipette, individual QC certificate and operating instructions, pipette charging stand, fast charger with AC adapter and cord, additional battery pack, Qualitips pipette tip samples.
Pipetting volume:
|
variable
|
Type (pipettes ...):
|
electronically controled pipettemultichannel pipette
|
Shipping weight: |
4,00 Kg |
|
Inaccuracy (Rel. err.%) |
Imprecision (CV %) |
|
Vol. µl |
Div. µl |
Chan. |
Min. vol. |
Mid. vol. |
Max. vol. |
Min. vol. |
Mid. vol. |
Max. vol. |
Tip style |
0.5 - 10 |
0.05 |
8 |
<± 3.5%** |
<± 1.5% |
<± 1.0% |
< 3.0%** |
< 0.9% |
< 0.7% |
ultra 10 µL |
2.5 - 50 |
0.25 |
8 |
<± 1.0%** |
<± 0.9% |
<± 0.8% |
< 1.0%** |
< 0.6% |
< 0.4% |
200 µL |
10 - 200 |
1.0 |
8 |
<± 0.9%** |
<± 0.7% |
<± 0.6% |
< 0.6%** |
< 0.4% |
< 0.25% |
200 µL |
20 - 350 |
5.0 |
8 |
<± 1.0%** |
<± 0.8% |
<± 0.6% |
< 0.6%** |
< 0.4% |
< 0.25% |
350 µL |
0.5 - 10 |
0.05 |
12 |
<± 3.5%** |
<± 1.5% |
<± 1.0% |
< 3.0%** |
< 0.9% |
< 0.7% |
ultra 10 µL |
2.5 - 50 |
0.25 |
12 |
<± 1.0%** |
<± 0.9% |
<± 0.8% |
< 1.0%** |
< 0.6% |
< 0.4% |
200 µL |
10 - 200 |
1.0 |
12 |
<± 0.9%** |
<± 0.7% |
<± 0.6% |
< 0.6%** |
< 0.4% |
< 0.25% |
200 µL |
20 - 350 |
5.0 |
12 |
<± 1.0%** |
<± 0.8% |
<± 0.6% |
< 0.6%** |
< 0.4% |
< 0.25% |
350 µL |
Performance measured at ** 10% of nominal volume