Julabo Temperature control system |
PRESTO W85 |
PRESTO W85t |
Working temperature range (°C) |
-85 to +250 |
-85 to +250 |
Temperature control |
ICC |
ICC |
Temperature stability (°C) |
+0.05 to +0.1 |
+0.05 to +0.1 |
Setting / display resolution (°C) |
+0.01 |
+0.01 |
Integrated programmer |
8x60 steps |
8x60 steps |
Temperature Display |
TFT Touchscreen |
TFT Touchscreen |
Heating capacity (kW) |
6 at 400V; 6 at 230V; 5.5 at 208V |
15 at 400V; 15 at 230V; 12.5 at 208V |
Cooling capacity (Medium: JULABO Thermal | Ethanol)
|
200 °C | 2.8 kW 20 °C | 2.5 kW 0 °C | 2.4 kW -20 °C | 2.4 kW -40 °C | 2.4 kW -60 °C | 2.2 kW -80 °C | 0.4 kW |
200 °C | 2.8 kW 20 °C | 2.5 kW 0 °C | 2.4 kW -20 °C | 2.4 kW -40 °C | 2.4 kW -60 °C | 2.2 kW -80 °C | 0.4 kW |
Pump capacity flow rate (l/min) |
35 to 80 |
35 to 80 |
Pump capacity flow pressure (bar)
|
0.48 to 3.2 at 400V 0.48 to 3.2 at 230V 0.48 to 3.0 at 208V |
0.48 to 3.2 at 400V 0.48 to 3.2 at 230V 0.48 at 3.0 at 208V |
Viscosity max. (cSt) |
50 |
50 |
Pump connections |
M30x1.5 |
M30x1.5 |
Refrigerant stage 1 |
R507 |
R507 |
Filling volume refrigerant stage 1 (g) |
700 |
700 |
Global Warming Potential for R507 |
3985 |
3985 |
Carbon dioxide equivalent stage 1 (t) |
2.79 |
2.79 |
Refrigerant stage 2 |
R23 |
R23 |
Filling volume refrigerant stage 2 (g) |
680 |
680 |
Global Warming Potential for R23 |
14800 |
14800 |
Carbon dioxide equivalent stage 2 (t) |
10.064 |
10.064 |
External Pt100 sensor connection |
integrated |
integrated |
Digital interfaces
|
RS232 SD memory card USB Ethernet Modbus Alarm-out Optional: RS485, Profibus |
RS232 SD memory card USB Ethernet Modbus Alarm-out Optional: RS485, Profibus |
Analog connection input / output |
Optional |
Optional |
Ambient temperature (°C) |
5 to 40 |
5 to 40 |
Dimensions W x L x H (cm) |
61 x 84.5 x 125 |
61 x 84.5 x 125 |
Weight (kg) |
335 |
335 |
Sound pressure level (distance 1 m) max. (dBA) |
69 |
69 |
Process volume min. (active heat exchanger volume) liters |
9.5 (5) |
9.5 (5) |
Internal usable expansion vol. (liters) |
7 |
7 |
Classification according to DIN12876-1 |
Classification III (FL) |
Classification III (FL) |
Cooling of compressor |
2-stage Water |
2-stage Water |
Cooling water connection |
G 3/4'' male with barbed fittings for tubing 1/2'' ID |
G 3/4'' male with barbed fittings for tubing 1/2'' ID |
Cooling water consumption (l/min) |
2 to 6 |
2 to 6 |
Cooling water temperature (°C) |
< 30 |
< 30 |
Cooling water differential pressure (bar) |
0.5 |
0.5 |
Available voltage versions
|
3 x 208-230V/60Hz (+/- 5%) / 30A at 208V, 33A at 230V / without Plug 3 x 230V/50Hz (+/- 10%) / 30A / Plug 63A CEE 3 x 400V/50Hz (+/- 10%) / 18A / Plug 32A CEE |
3 x 208-230V/60Hz (+/- 5%) / 52A at 208V, 55A at 230V / without Plug 3 x 230V/50Hz (+/- 10%) / 52A / Plug 63A CEE 3 x 400V/50Hz (+/- 10%) / 31A / Plug 32A CEE |