ELGA BIOPURE 120. Order online at ProfiLab24 .com with top advice and 24/7 service. Fast delivery. Secure payment.
ELGA BIOPURE 120The BIOPURE 60/120 has an in-built reservoir to ensure minimal space is occupied in the hospital or surgery. Contamination risk to patients is minimized by the use of compliant water maintained through the use of in-line UV oxidation, deionization and molecular ultrafiltration.
Features
- Delivers up to 10 liters per minute of compliant water.Meets the latest medical standards (EN15883, HTM2030/2031, NHS MESc52, MDA/HIS, ISO 15883 part 4, EN285)
- Guaranteed bacterial specification of <1 CFU/10ml provided by the integrated purification technologies all located inside one single system
- Easy to use monthly disinfection program
- Easy to install under a bench
- Full service and validation support
To operate a BIOPURE system is a starter kit 1 or 2 is required.
Booster pump:
|
without
|
Power:
|
120 l/h
|
Modules:
|
UV-modulesRO-moduleultrafiltration module
|
Storage tank - tank:
|
with
|
Water treatment type:
|
pure water system
|
Water source:
|
from the water mains
|
Product weight: |
58,00 Kg |
MODEL |
BIOPURE 120 |
Treated Water Specifications |
|
Make-up rate at 15°C |
120 l/hr |
Delivery flow rate - maximum |
Up to 10 l/min at 22.5psi (1.5bar) |
Inorganics (conductivity at 25°C) |
<30 µS/cm1)
|
Inorganics (conductivity at 25°C) (Optipure Purification Cartridge installed) |
<5 µS/cm1)
|
Bacteria - typical |
<1 CFU/10ml2)
|
Endotoxin |
<0.25 EU/ml |
Dimensions and weights |
|
Dimensions |
Height 820mm (32.8in) 834mm (33.4in) including castors Width 790mm (31.1in) Depth 480mm (18.9in) |
Supply weight |
58 kg (127.7lb) |
Operational weight |
115 kg (253.5lb) |
Installation |
Floor mounted3)
|
Feedwater Requirement |
|
Source |
Tap water as detailed below |
Conductivity |
<2000 µS/cm *
|
Contaminant |
|
Hardness |
<250 ppm as CaCO3
|
Free Chlorine |
<1 ppm Cl2
|
Chloramine |
<0.2 ppm Cl2
|
Silica |
<30 ppm SiO2
|
Fouling Index |
<10 FI |
Iron/manganese |
<0.05Fe/Mn ppm |
Organics |
<3 ppm TOC |
Temperature |
4 - 40°C (Recommended 15 - 25°C) |
Flowrate (requirement at 15°C) |
9 l/min |
Drain requirements (gravity fall with air gap) |
20 l/min |
Feedwater pressure |
90psi (6bar) maximum 30psi (2bar) minimum |
Electrical Requirements |
|
Mains Input |
230V ac, 50Hz |
System control voltage (not including pumps and UV) |
24V dc |
Power consumption (peak demand) |
800VA |
Electrical protection rating |
13 amp supply with suitable ELB |
Noise level |
<60 dBA |
1)Subject to a maximum feed water TDS of 500 ppm. For applications outside of this range, please contact Technical Support.
2)System to be installed and regularly disinfected following ELGA LabWater installation design guidelines. Sample to be taken from hygienic sample valve and sampled in accordance with ELGA LabWater Guidelines.
3)Install in a properly ventilated area.
* Deionization cartridge life may vary with feedwaters >1400 µS/cm
Failure to comply with the feed water pretreatment recommendations will affect the life and performance of key components and may invalidate the warranty. Feedwater requirements are specified for systems fitted with a Protek pretreatment cartridge. If feedwater purity is variable or values are outside any of the above ranges, additional pretreatment is recommended to be installed in the feedwater supply to the BIOPURE system.