The BIOPURE 60/120 has an in-built reservoir to ensure minimal space is occupied in the hospital or surgery. Contamination risk to patients is minimized by the use of compliant water maintained through the use of in-line UV oxidation, deionization and molecular ultrafiltration.

Features
  • Delivers up to 10 liters per minute of compliant water.Meets the latest medical standards (EN15883, HTM2030/2031, NHS MESc52, MDA/HIS, ISO 15883 part 4, EN285)
  • Guaranteed bacterial specification of <1 CFU/10ml provided by the integrated purification technologies all located inside one single system
  • Easy to use monthly disinfection program
  • Easy to install under a bench
  • Full service and validation support


To operate a BIOPURE system is a starter kit 1 or 2 is required.

Booster pump: without
Power: 120 l/h
Modules: UV-modulesRO-moduleultrafiltration module
Storage tank - tank: with
Water treatment type: pure water system
Water source: from the water mains
Product weight: 58,00 Kg
MODEL BIOPURE 120
Treated Water Specifications
Make-up rate at 15°C 120 l/hr
Delivery flow rate - maximum Up to 10 l/min at 22.5psi (1.5bar)
Inorganics (conductivity at 25°C) <30 µS/cm1)
Inorganics (conductivity at 25°C)
(Optipure Purification Cartridge installed)
<5 µS/cm1)
Bacteria - typical <1 CFU/10ml2)
Endotoxin <0.25 EU/ml
Dimensions and weights
Dimensions Height 820mm (32.8in)
834mm (33.4in) including castors
Width 790mm (31.1in)
Depth 480mm (18.9in)
Supply weight 58 kg (127.7lb)
Operational weight 115 kg (253.5lb)
Installation Floor mounted3)
Feedwater Requirement
Source Tap water as detailed below
Conductivity <2000 µS/cm *
Contaminant
Hardness <250 ppm as CaCO3
Free Chlorine <1 ppm Cl2
Chloramine <0.2 ppm Cl2
Silica <30 ppm SiO2
Fouling Index <10 FI
Iron/manganese <0.05Fe/Mn ppm
Organics <3 ppm TOC
Temperature 4 - 40°C
(Recommended 15 - 25°C)
Flowrate (requirement at 15°C) 9 l/min
Drain requirements (gravity fall with air gap) 20 l/min
Feedwater pressure 90psi (6bar) maximum
30psi (2bar) minimum
Electrical Requirements
Mains Input 230V ac, 50Hz
System control voltage
(not including pumps and UV)
24V dc
Power consumption (peak demand) 800VA
Electrical protection rating 13 amp supply with suitable ELB
Noise level <60 dBA


1)Subject to a maximum feed water TDS of 500 ppm. For applications outside of this range, please contact Technical Support.
2)System to be installed and regularly disinfected following ELGA LabWater installation design guidelines. Sample to be taken from hygienic sample valve and sampled in accordance with ELGA LabWater Guidelines.
3)Install in a properly ventilated area.
* Deionization cartridge life may vary with feedwaters >1400 µS/cm

Failure to comply with the feed water pretreatment recommendations will affect the life and performance of key components and may invalidate the warranty. Feedwater requirements are specified for systems fitted with a Protek pretreatment cartridge. If feedwater purity is variable or values are outside any of the above ranges, additional pretreatment is recommended to be installed in the feedwater supply to the BIOPURE system.