ELGA BIOPURE 15The BIOPURE 7/15 is a compact system able to fit under or on a bench or wall mounted. Utilizing integrated purification technologies it delivers purified water at the point of use for washer disinfectors and autoclaves to minimize the contamination risk to patients.
Features
- Delivers up to 7 liters an hour of compliant water
- Meets the latest medical standards (EN15883, HTM2030/2031, NHS MESc32, MDA/HIS, ISO 15883 part 4, EN285)
- Low running costs are comparable to bottled water costs
- Guaranteed bacterial specification, provided by integrated purification technologies
- Compact size makes it suitable to be installed under a bench, mounted on a wall or fitted into a cabinet
- Full service and validation support
To operate a system BIOPURE a tank with 5 level switches is required.
Booster pump:
|
withoutwith
|
Power:
|
15 l/h
|
Storage tank - tank:
|
without
|
Water treatment type:
|
pure water system
|
Water source:
|
from the water mains
|
Shipping weight: |
25,00 Kg |
Product weight: |
21,00 Kg |
MODEL |
BIOPURE 15 |
Treated Water Specifications |
|
Make-up rate at15°C |
15 l/hr |
Delivery flow rate - maximum |
2 l/min at 15psi (1bar) |
Inorganics (conductivity at 25°C) |
<1 µS/cm |
Bacteria - typical |
<1 CFU/10ml1)
|
Endotoxin |
<0.25 EU/ml |
Dimensions and weights |
|
Dimensions |
Height 460mm (18.1in) Width 550mm (21.7in) Depth 270mm (10.6in) |
Operational weight |
21Kg |
Installation |
Bench/Wall2)
|
Feedwater Requirement |
|
Source |
Tap water as detailed below |
Conductivity |
<1400 µS/cm |
Contaminant |
|
Hardness |
<250 ppm as CaCO3
|
Free chlorine |
0.5 ppm |
Chloramine |
<0.2 ppm |
Silica |
<30 SiO2 ppm |
Fouling index |
<10 FI |
Iron/manganese |
<0.2 ppm |
Organics |
N/A |
Temperature |
4 - 40°C (Recommended 15 - 25°C) |
Flowrate (requirement at 15°C) |
1.4 l/min |
Drain requirements (gravity fall with air gap) |
1.2 l/min |
Feedwater Pressure |
|
Maximum without internal boost pump |
90psi (6.0bar) |
Minimum without internal boost pump |
60psi (4.0bar) |
Maximum with internal boost pump |
30psi (2.0bar) |
Maximum without internal boost pump |
Flooded Suction |
Electrical Requirements |
|
Mains Input |
100-240V ac, 50-60Hz |
System control voltage (not including pumps and UV) |
24V dc |
Power consumption with boost pump |
80VA |
Power consumption without boost pump |
50VA |
Electrical protection rating |
13 amp supply with suitable ELB |
Noise level |
<45 dBA |
1)System to be installed and regularly disinfected following ELGA LabWater installation design guidelines. Sample to be taken from hygienic sample valve and sampled in accordance with ELGA LabWater guidelines.
2)Install in a properly ventilated area.