ELGA BIOPURE 600

The BIOPURE 200/300/600 reduces the contamination risk to patients by feeding high purity water at the point of use to large automated washer disinfectors and endoscope reprocessors. The quality of the purified water is maintained through the use of in-line UV sterilization, ion exchange and molecular ultrafiltration.

Features
  • Delivers up to 30 liters per minute of compliant water
  • Meets the latest medical standards (EN15883, HTM2030/2031, NHS MESc52, MDA/HIS, ISO 15883 part 4, EN285)
  • Guaranteed bacterial specification of <1 CFU/10ml provided by the integrated purification technologies all located inside one single system.
  • Easy to use monthly disinfection program.
  • Compact size due to its integrated 350 liter reservoir
  • Full service and validation support

Modell: BIOPURE 600
Water treatment model: pure water system
Booster pump: without
Capacity: 600 l/h
Storage tank - tank: with
Water source: from the water mains
Modules: UV-modules RO-module
Product weight:210,00 Kg
MODEL

BIOPURE 600

Treated Water Specifications

 

Make-up rate at 15°C

600 l/hr

Delivery flow rate - maximum

Up to 30 l/min at 52psi (3.5bar)

Inorganics (conductivity at 25°C)

<30µS/cm

Bacteria - typical

<1CFU/10ml1)

Endotoxin

<0.25 EU/ml

Dimensions and weights

 

Dimensions

Height 1820 mm (71.7in)
Width 756 mm (28.75in)
Depth 1100 mm (35in)

Supply weight

210 kg (463 lb)

Operational weight

560 kg (1235 lb)

Installation

Floor Mounted2)

Electrical Requirements

 

Mains Input

230V ac, 50Hz/60Hz

System control voltage
(not including pumps and UV)

24V dc

Power consumption (peak demand)

2200VA

Electrical protection rating

20 amps

Noise level during recirculation

<70dBA

Feedwater Requirement

 

Source

Tap water as detailed below

Conductivity

<1400 µS/cm

Temperature

4 - 40°C (Recommended 15 - 25°C)

Flowrate (requirement at 15°C)

20 l/min

Drain requirements
(gravity fall with air gap)

45 l/min

Feedwater pressure

150psi (10bar) maximum
30psi (2bar) minimum

Contaminant

 

Measure Range Pretreatment
CalciumCa ppm as CaCO3<250None3)
<250Softner or use very low RO recovery3)
Free chlorineCl2 ppm<0.1None
0.1 - 0.520 inch carbon block
>0.5Cylinder of carbon sized correctly to obtain <0.1ppm
ChloramineCl2 ppm<0.1None
0.1 - 0.520 inch high efficiency carbon4)
>0.5Cylinder of carbon sized correctly to obtain <0.1ppm
SilicaSiO2 ppm<30None
>3020 inch cartridge depth filter or use very low RO recovery4)
Fouling indexFI<10None
10 to 2020 inch cartridge depth filter
>20Back washable media filter with a minimum flow rate of 20 l/min
Iron/manganeseFe/Mn ppm<0.05None
>0.0520 inch cartridge depth filter5)
>0.1Back-washable Fe filter6)
Organics TOCppm C<2None
2 to 320 inch carbon block4)
>3Cylinder of carbon sized correctly for TOC demand4)


1)System to be installed and regularly disinfected following ELGA LabWater installation design guidelines. Sample to be taken from hygienic sample valve and sampled in accordance with ELGA LabWater Guidelines.
2)Install in a properly ventilated area. All connections - Inlet, sanitization drain, reservoir drain, ringmain outlet and return are 3/4" BSPP (female).
3)Check LSI, increase frequency of acid cleaning.
4)Increase frequency of alkaline cleaning.
5)Please use ELGA Part Numbers LA683 (Filter Housing) and LC172 (Carbon Filter).