ELGA BIOPURE 600The BIOPURE 200/300/600 reduces the contamination risk to patients by feeding high purity water at the point of use to large automated washer disinfectors and endoscope reprocessors. The quality of the purified water is maintained through the use of in-line UV sterilization, ion exchange and molecular ultrafiltration.
Features
- Delivers up to 30 liters per minute of compliant water
- Meets the latest medical standards (EN15883, HTM2030/2031, NHS MESc52, MDA/HIS, ISO 15883 part 4, EN285)
- Guaranteed bacterial specification of <1 CFU/10ml provided by the integrated purification technologies all located inside one single system.
- Easy to use monthly disinfection program.
- Compact size due to its integrated 350 liter reservoir
- Full service and validation support
Booster pump:
|
without
|
Power:
|
600 l/h
|
Modules:
|
RO-moduleUV-modules
|
Storage tank - tank:
|
with
|
Water treatment type:
|
pure water system
|
Water source:
|
from the water mains
|
Shipping weight: |
210,00 Kg |
Product weight: |
210,00 Kg |
MODEL |
BIOPURE 600
|
Treated Water Specifications |
|
Make-up rate at 15°C |
600 l/hr
|
Delivery flow rate - maximum |
Up to 30 l/min at 52psi (3.5bar)
|
Inorganics (conductivity at 25°C) |
<30µS/cm
|
Bacteria - typical |
<1CFU/10ml1)
|
Endotoxin |
<0.25 EU/ml
|
Dimensions and weights |
|
Dimensions |
Height 1820 mm (71.7in) Width 756 mm (28.75in) Depth 1100 mm (35in)
|
Supply weight |
210 kg (463 lb)
|
Operational weight |
560 kg (1235 lb)
|
Installation |
Floor Mounted2)
|
Electrical Requirements |
|
Mains Input |
230V ac, 50Hz/60Hz
|
System control voltage (not including pumps and UV) |
24V dc
|
Power consumption (peak demand) |
2200VA
|
Electrical protection rating |
20 amps
|
Noise level during recirculation |
<70dBA
|
Feedwater Requirement |
|
Source |
Tap water as detailed below
|
Conductivity |
<1400 µS/cm
|
Temperature |
4 - 40°C (Recommended 15 - 25°C)
|
Flowrate (requirement at 15°C) |
20 l/min
|
Drain requirements (gravity fall with air gap) |
45 l/min
|
Feedwater pressure |
150psi (10bar) maximum 30psi (2bar) minimum
|
Contaminant |
|
|
Measure |
Range |
Pretreatment |
Calcium |
Ca ppm as CaCO3
|
<250 |
None3) |
|
|
<250 |
Softner or use very low RO recovery3)
|
Free chlorine |
Cl2 ppm |
<0.1 |
None |
|
|
0.1 - 0.5 |
20 inch carbon block |
|
|
>0.5 |
Cylinder of carbon sized correctly to obtain <0.1ppm |
Chloramine |
Cl2 ppm |
<0.1 |
None |
|
|
0.1 - 0.5 |
20 inch high efficiency carbon4) |
|
|
>0.5 |
Cylinder of carbon sized correctly to obtain <0.1ppm |
Silica |
SiO2 ppm |
<30 |
None |
|
|
>30 |
20 inch cartridge depth filter or use very low RO recovery4)
|
Fouling index |
FI |
<10 |
None |
|
|
10 to 20 |
20 inch cartridge depth filter |
|
|
>20 |
Back washable media filter with a minimum flow rate of 20 l/min |
Iron/manganese |
Fe/Mn ppm |
<0.05 |
None |
|
|
>0.05 |
20 inch cartridge depth filter5)
|
|
|
>0.1 |
Back-washable Fe filter6)
|
Organics TOC |
ppm C |
<2 |
None |
|
|
2 to 3 |
20 inch carbon block4)
|
|
|
>3 |
Cylinder of carbon sized correctly for TOC demand4)
|
1)System to be installed and regularly disinfected following ELGA LabWater installation design guidelines. Sample to be taken from hygienic sample valve and sampled in accordance with ELGA LabWater Guidelines.
2)Install in a properly ventilated area. All connections - Inlet, sanitization drain, reservoir drain, ringmain outlet and return are 3/4" BSPP (female).
3)Check LSI, increase frequency of acid cleaning.
4)Increase frequency of alkaline cleaning.
5)Please use ELGA Part Numbers LA683 (Filter Housing) and LC172 (Carbon Filter).
6)Increase frequency of acid cleaning.