ELGA MEDICA Pro RE30

Item number: M_ELGA_21054

EAN: 4064343193436

Category: ELGA

Manufacturer: VWS Deutschland

Price on application

Product price on request.


ELGA MEDICA Pro RE30

The MEDICA Pro supplies Clinical Laboratory Reagent Water (CLRW) standard (formally CLSI Type 1) water using UV and microfiltration technologies together with recirculation and ion exchange. This compact system is ideal to directly feed large or linked multiple analyzers.

Features
  • Delivers up to 4 liters per minute of CLRW standard water (formally CLSI Type 1 water)
  • Bacterial specification <1CFU/ml. Simple sanitization programme runs quickly
  • Compatible with single and multiple clinical diagnostic analyzers which require up to 4 liters per minute of CLRW standard water
  • Low running and consumable costs
  • Space saving design with its integrated wrap around 50 liter reservoir
  • A bypass loop for continual supply of compliant water in emergency situations to the analyzer
  • Easy to install under a bench
  • Full service and validation support


For operating a MEDICA Pro system, a starter kit 1, 2 or 3 is required.

Booster pump: without
Power: 30 l/h
Storage tank - tank: with
Water treatment type: ultrapure water system
Water source: from the water mains
Product weight: 53,00 Kg
MODEL MEDICA Pro-RE30
Treated Water Specifications
Make-up rate at 15°C 30 l/hr
Delivery flow rate - maximum Up to 4 l/min at 14psi (1bar) 1)
Inorganics (resistivity at 25°C) >10 MOhm-cm
Organics (TOC) - typical <30 ppb
Bacteria - typical <1 CFU/ml 2)
Particles 0.05 µm filter
Silica <0.05 mg/l
Dimensions and weights
Dimensions Height 820mm (32.8in)
Height 834mm (33.4in) including castors
Width 794mm (31.8in)
Depth 470mm (18.8in)
Supply weight 45kg
Operational weight 102Kg
Installation Floor/bench
Feedwater Requirement
Source Tap water as detailed below
Conductivity <2000 µS/cm 3)
Contaminant
Hardness <250 ppm as CaCO3
Free Chlorine <4 ppm Cl2
Chloramine <1 ppm Cl2
Silica <30 ppm SiO2
Fouling Index <10 FI
Iron/manganese <0.1 ppm
Organics <3 ppm TOC
Temperature 4 - 40°C
(Recommended 15 - 25°C)
Flowrate
(requirement at 15°C)
9 l/min
Drain requirements
(gravity fall with air gap)
20 l/min
Feedwater pressure 90psi (6bar) maximum
30psi (2bar) minimum
Electrical Requirements
Mains Input 230V ac, 50Hz
115V, 60Hz
System control voltage
(not including pumps and UV)
24V dc
Power consumption
(peak demand)
650VA
Electrical protection rating 10 amps
Noise level <63dBA


1) With high flow kit (LA700) or 2 l/min at 14 psi (1bar) without high flow kit
2) Subject to correct operating and maintenance procedures
3) Purification pack life may vary with feedwater >1400 µS/cm