ELGA MEDICA R 15

The MEDICA 7/15 supplies a constant flow of high purity water which is maintained by recirculation through ion exchange resins, UV and an integral 0.2µm microfilter. It is an ideal choice for use with a single chemistry or immunoassay clinical analyzer.

Features
  • Delivers up to 1.8 liters per minute of CLRW standard water (formally CLSI Type 1 water)
  • Provides compliant purified water suitable for a dissolved oxygen specification
  • Bacterial specification <1CFU/ml.Compatible with single clinical diagnostic analyzers which require up to 1.8 liters per minute of CLRW standard water
  • A bypass loop for continual supply of compliant water in emergency situations to the analyzer
  • Low running and consumable costs
  • Compact size makes it suitable to be installed under a bench or mounted on a wall
  • Full service and validation support


To operate a system MEDICA a tank with 5 level switches is required.

Modell: R 15 Standard
Water treatment model: pure water system
Booster pump: with without
Capacity: 15 l/h
Storage tank - tank: without
Water source: of ion exchanger from reverse osmosis system from the water mains
Shipping weight:24,00 Kg
Product weight:19,00 Kg
MODEL MEDICA-R 15
Treated Water Specifications
Make-up rate at 15°C15 l/hr
Delivery flow rate - maximumUp to 1.8 l/min at 30psi (2bar)
Inorganics (resistivity at 25°C)>10 MOhm-cm
Organics (TOC) - typical<30 ppb
Bacteria - typical<1 CFU/ml
Particles0.2 µm filter
Silica<0.05 mg/l
Dissolved oxygen (MEDICA-D)N/A
Dimensions and weights
Dimensions Height 460 mm (18.1in)
Width 550 mm (21.7in)
Depth 270mm (10.6in)
Operational weight19Kg
InstallationBench/wall
Feedwater Requirement
SourceTap water as detailed below
Conductivity<1400 µS/cm
Contaminant
Hardness<350 ppm as CaCO3
Free chlorine<0.5 ppm
Chloramine<0.2 ppm
Silica<30 ppm
Fouling index<10
Iron/manganese<0.2 ppm
Organics<3 ppm TOC
Temperature4 - 40°C
(Recommended 15 - 25°C)
Flowrate (requirement at 15°C)1.4 l/min
Drain requirements (gravity fall with air gap)1.2 l/min
Feedwater Pressure
Maximum without internal boost pump90psi (6bar)
Minimum without internal boost pump60psi (4bar)
Maximum with internal boost pump30psi (2bar)
Minimum with internal boost pumpFlooded Suction
Electrical Requirements
Mains Input100-240V ac, 50-60Hz
System control voltage
(not including pumps and UV)
24V dc
Power consumption with boost pump83VA
Power consumption without boost pump60VA
Noise level<45dBA