ELGA PURELAB 3120The PURELAB 3000 series is ideal for all high usage applications requiring Type III water. The flexibility of the system means that it can be easily upgraded to provide Type II pure water or increased capacity should requirements change. Pure water can be used direct from the internal 50 liter reservoir or fed to much larger vessels as required.
- Auto rinse and simple sanitization procedures maintain the purity of water for the life of the system
- A variety of connections and distribution pumps are available to cater for a diverse range of applications
For operating a PURELAB 3000/7000 system, a starter kit 1 or 2 is required.
Booster pump:
|
without
|
Power:
|
120 l/h
|
Storage tank - tank:
|
with
|
Water treatment type:
|
pure water system
|
Water source:
|
of ion exchangerfrom the water mains
|
Product weight: |
60,00 Kg |
Model |
PURELAB 3120 |
Treated Water Specifications |
|
Make-up rate at 15°C |
120 l/hr |
Daily usage - typical |
960 liters |
Daily usage - maximum |
2880 liters |
Delivery flowrate |
- |
Inorganics rejection |
>95% |
Inorganics (resistivity at 25°C) |
>10 MOhm-cm 1)2) |
Bacteria - typical |
- |
Organics (TOC) - typical |
<50 ppb 1)
|
Particles |
>99% Rejection 1)
|
Dimensions and weights |
|
Dimensions |
Height 820mm (32.8in) 834mm (33.4in) including castors Width 794mm (31.8in) Depth 470mm (18.8in) |
Supply weight |
60kg |
Operational weight |
117kg |
Installation |
Floor/bench |
Feedwater Requirements |
|
Source |
Tap water as detailed below |
Contaminant |
|
Conductivity |
<2000 µS/cm*
|
Hardness |
<250 ppm as CaCO3
|
Free chlorine |
<4 ppm Cl2
|
Chloramine |
<1 ppm as Cl2
|
Silica |
<30 ppm SiO2
|
Fouling Index |
<10 FI |
Iron/manganese |
<0.1 ppm Fe/Mn |
Organics |
<3 ppm TOC |
Temperature |
4 - 40°C (Recommended 15 - 25°C) |
Flowrate (maximum requirement at 15°C) |
9 l/min |
Drain requirements (gravity fall with air gap) |
20 l/min |
Feedwater pressure |
6 bar (90 psi) maximum 2 bar (30 psi) minimum |
Electrical Requirements |
|
Mains input |
230V ac, 50Hz 115V ac, 60Hz |
System control voltage (not including pumps and UV) |
24V dc |
Power consumption (peak demand) |
650VA |
Electrical protection rating |
10 amps |
Specifications are for a system fed with a suitable water supply and installed, operated and sanitized according to the operator manuals.
1) Quality of water feeding t he reservoir. Quality out of reservoir depends on use and design.
2) With optional deionization cartridge fitted.
3) With High Flow Kit (LA723) or 2 l/min at 12 psi (0.8 bar) without high flow kit
Failure to comply with the feedwater pretreatment recommendations will affect the life and performance of key components and may invalidate the warranty. Feedwater requirements are specified for units fitted with a Protek L2 pretreatment cartridge. If feedwater purity is variable or values are outside any of the above ranges, additional pretreatment is recommended to be installed in the feedwater supply to the unit.
* Deionization cartridge life may vary with feedwater >1400 µS/cm