ELGA PURELAB 7120

Item number: ELGA_21046>

EAN: 4064343194396

Category: ELGA

Manufacturer: VWS Deutschland

Price on application

Product price on request.


ELGA PURELAB 7120

The PURELAB 7000 series is designed for applications requiring higher flows of Type II pure water or as a pure water feed system for small distribution loops. The system can supply water from the internal 50 liter reservoir at varying flow rates or can feed into much larger vessels as required.
  • Small high purity distribution system with variable flow rates
  • Auto rinse and simple sanitization procedures maintain the purity of water for the life of the system


For operating a PURELAB 3000/7000 system, a starter kit 1 or 2 is required.

Booster pump: without
Power: 120 l/h
Storage tank - tank: with
Water treatment type: pure water system
Water source: of ion exchangerfrom the water mains
Product weight: 60,00 Kg
Model PURELAB 7120
Treated Water Specifications
Make-up rate at 15°C 120 l/hr
Daily usage - typical 960 liters
Daily usage - maximum 2880 liters
Delivery flowrate Up to 4 l/min at 12 psi (0.8 bar) 1)
Inorganics rejection -
Inorganics (resistivity at 25°C) >15 MOhm-cm
Bacteria - typical <1 CFU/ml
Organics (TOC) - typical <10 ppb
Particles 0.05 µm
Dimensions and weights
Dimensions Height 820mm (32.8in)
834mm (33.4in) including castors
Width 794mm (31.8in)
Depth 470mm (18.8in)
Supply weight 53kg
Operational weight 110kg
Installation Floor/bench
Feedwater Requirements
Source Tap water as detailed below
Contaminant
Conductivity <2000 µS/cm*
Hardness <250 ppm as CaCO3
Free chlorine <4 ppm Cl2
Chloramine <1 ppm as Cl2
Silica <30 ppm SiO2
Fouling Index <10 FI
Iron/manganese <0.1 ppm Fe/Mn
Organics <3 ppm TOC
Temperature 4 - 40°C
(Recommended 15 - 25°C)
Flowrate (maximum requirement at 15°C) 9 l/min
Drain requirements (gravity fall with air gap) 20 l/min
Feedwater pressure 6 bar (90 psi) maximum
2 bar (30 psi) minimum
Electrical Requirements
Mains input 230V ac, 50Hz
115V ac, 60Hz
System control voltage (not including pumps and UV) 24V dc
Power consumption (peak demand) 650VA
Electrical protection rating 10 amps


Specifications are for a system fed with a suitable water supply and installed, operated and sanitized according to the operator manuals.
1) With High Flow Kit (LA723) or 2 l/min at 12 psi (0.8 bar) without high flow kit

Failure to comply with the feedwater pretreatment recommendations will affect the life and performance of key components and may invalidate the warranty. Feedwater requirements are specified for units fitted with a Protek L2 pretreatment cartridge. If feedwater purity is variable or values are outside any of the above ranges, additional pretreatment is recommended to be installed in the feedwater supply to the unit.

*
Deionization cartridge life may vary with feedwater >1400 µS/cm