ELGA PURELAB flex 2The PURELAB flex product range
Security for analytical & life science applications
- Reliable ultrapure water up to 18.2 megohm cm
- Real-time monitoring of the TOC content
- Clear display shows all important information
- Data acquisition via USB interface for quality assurance
- Quick and easy disinfection
Designed for laboratory needs for today
- intuitive operation
- Ergonomic Multi Function Dispenser
- Height adjustable and flexible takeout
- Table or wall installation possible
- Flexible water supply (including variable flow up to 2 l / min and automatic extraction of volume to 60 liters)
To operate a PURELAB flex 2 system is a purification cartridge type LC208 is required, it must be ordered separately.
Booster pump:
|
without
|
Power:
|
2 l/min
|
Modules:
|
UV-modules
|
Storage tank - tank:
|
without
|
Water treatment type:
|
ultrapure water system
|
Water source:
|
of ion exchangerfrom reverse osmosis system
|
Shipping weight: |
16,00 Kg |
Product weight: |
11,00 Kg |
model |
PURELAB flex 2 |
Specifications Product water |
|
Daily water quantity |
> 10 liters |
Max. Flow rate |
up to 2 l/min |
inorganic substances at 25°C |
18,2 MOhm-cm |
organic substances (TOC) - typical |
< 5ppb 1)
|
Directly from the integrated tank |
- |
Bacteria - typical (with final filter) |
< 1 KBE/10 ml |
Bacteria - typical (with biofilters) |
< 1 KBE/10 ml |
Endotoxin (with biofilters) |
< 0,001 EU/ml |
DNase (with biofilters) |
< 20 pg/ml |
RNase (with biofilters) |
< 0,002 ng/ml |
Dimensions and weights |
|
dimensions |
width 236 mm depth 374 mm min. height 900 mm max. height 1020 mm |
operating weight |
11 kg (24,2 lb) |
installation |
Desk / Wall |
Specifications feedwater |
|
source of supply |
Drinking water with subsequent treatment, preferably by reverse osmosis (RO) or ion exchanger (SDI) - filtered or distilled |
conductivity |
< 1 µS/cm 2) |
ingredients |
|
hardness |
- |
Free chlorine |
< 0,05 ppm Cl2
|
chloramine |
< 0,05 ppm Cl2
|
chlorine Total |
< 0,05 ppm Cl2 |
silicate |
< 2 ppm |
carbon dioxide maximum |
< 0,1 ppm |
Fouling Index |
< 1 |
Iron / manganese |
- |
organic substances (TOC) |
- |
particle |
A 0.2 micron membrane prefilters is for all feedwater systems (except reverse osmosis) is recommended for longer tool life of final filters |
temperature |
4 - 40°C (recommended 10 - 25°C) |
flow (at 15°C) |
< 2 l/min |
Runoff Specification (free expiry) |
not required |
Feed water pressure |
|
maximum |
22 psi (1,5 bar) |
minimum |
minimum positive admission pressure |
Electrical specifications |
|
Mains power input |
100 - 240 V ac, 50 - 60 Hz |
System voltage (without pump and UV) |
24 V dc |
Energy consumption at peak load |
120 W |
noise level |
< 40 dBa |
1) Depending on the feedwater