ELGA PURELAB Option-S 7 BPThe PURELAB Option provides Type II +/II pure water on demand with a wide range of flow-rates to suit your needs. Applications range from glassware washing and the feeding of ultrapure water systems to cell culture and media preparation. The PURELAB Option-R provides additional bacteria and inorganic quality for your more critical applications.
- Delivers market leading microbial performance. In addition to simple sanitization and composite vent filtration, the PURELAB Option-R system is the only fully recirculating Type II+ pure water system on the market, ensuring high specification water
- Features "Reverse Osmosis feed optimized" resin mixes to increase the ion exchange capacity of consumables and minimize running costs
- Quick and easy sanitization and replacement of consumables to reduce maintenance time
- PURELAB Option systems are designed to be easy to access whether wall or bench mounted with a convenient dispense tap. They can be used with our wrap-around reservoirs to save space whilst optimizing purity
To operate a PURELAB Option-S / R Systems, a tank with 5 level switches is required.
Booster pump:
|
with
|
Power:
|
1,0 l/min
|
Modules:
|
RO-module
|
Storage tank - tank:
|
without
|
Water treatment type:
|
pure water system
|
Water source:
|
from the water mains
|
Shipping weight: |
21,00 Kg |
Product weight: |
16,00 Kg |
Model |
Option-S 7 |
Treated Water Specifications |
|
Make up rate 1) |
7,5 l/h |
Dispense rate from tap |
- |
Daily output (nominal max) 1) |
180 l/24 hour day |
Output reverse pressure (max.) 2) |
0,1 bar (1 psi) |
Purity |
|
Inorganics at 25°C |
1 to >10 MOhm-cm |
Total organic carbon (TOC) |
<30 ppb |
Bacteria |
- |
pH |
Effectively neutral |
Particles |
- |
Dimensions and weights |
|
Height |
460 mm (18,1") |
Width |
410 mm (16,1") |
Depth |
270 mm (10,6") |
Weight with internal boost pump (BP)
|
16 kg (35 lb) |
Weight without internal boost pump |
13,5 kg (30 lb) |
Feedwater Requirements |
|
Source Quality |
Potable mains water supply |
Fouling Index (max.) |
10 |
Total dissolved solids (max.) |
1400 µS/cm |
Free chlorine (max.) |
0,5 ppm |
Heavy metals (max.) |
0,05 ppm |
Silica (max.) |
30 ppm |
Temperature |
1 - 35°C |
Flowrate (maximum requirement) |
78 l/h |
Drain requirements (gravity fall with air gap). Maximum during service |
70 l/h |
Feedwater Pressure |
|
Maximum - without internal boost pump |
6,0 bar (90 psi) |
Minimum - without internal boost pump |
4,0 bar (60 psi) |
Maximum - with internal boost pump (BP)
|
2,0 bar (30 psi) |
Minimum - with internal boost pump (BP)
|
Flooded Suction |
Electrical Requirements |
|
Mains input |
100 - 240 V ac, 50 - 60 Hz |
System voltage |
24 V dc |
Power consumption with boost pump (BP)
|
43 VA |
Power consumption without boost pump |
19 VA |
Fuses |
2 x T 3,15A |
Reservoir level connection |
Jack Plug 3.5mm |
Noise level |
<45 dBA |
1 Standard conditions are 4 bar inlet pressure, 0 bar back pressure at 15 degrees centigrade, fed with potable water and a clean pre-treatment cartridge.
2Subject to suitable feedwater