ELGA PURELAB Ultra IonicThe new system is the intelligent PURELAB Ultra pure water system for the most critical applications. It differs from other systems through a variety of innovative and unique quality features and sets new standards in the production of ultrapure water quality of 18.2 MOhm-cm and higher.
- Maximum inorganic purity by unique PureSure system
- Real-time TOC monitoring provides reliable information on the organic quality in the two-second intervals
- Economic working at very high product water safety thanks PureSure system
- Constant low bacterial content through complete disinfection of all components in contact with water
To operate a PURELAB Ultra system two twin pack preparation cartridges Type LC182, 183, 184 o. 185 required.
The Twin Pack purification cartridges type LC182, 183, 184 o. 185 must be ordered separately.
Booster pump:
|
without
|
Power:
|
2 l/min
|
Modules:
|
UV-modules
|
Storage tank - tank:
|
without
|
Water treatment type:
|
ultrapure water system
|
Water source:
|
of ion exchangerfrom reverse osmosis system
|
Shipping weight: |
20,00 Kg |
Product weight: |
14,50 Kg |
model |
Ionic |
Treated Water Specifications |
|
Flowrate |
2,0 l/min max. |
Inorganics |
18,2 MOhm-cm |
TOC |
3 - 10 ppb 1)
|
Bacteria |
<0,1 KBE/ml 2) |
Bacterial endotoxin |
- |
pH |
Effectively neutral |
Particles |
0,05 µm |
RNase/DNase |
- |
Labpure cartridge capacity (LC 182) |
Liters at 18,2 MOhm-cm = 80,000 / µS/cm + ( 2,3 x ppm CO2 ) |
Dimension and weight |
|
Dimension |
Height 490 mm (19,3") Width 410 mm (16,2") Depth 365 mm (14,4") |
weight |
14,5 kg (32,0 lb) |
Feedwater Requirements |
|
Parameter |
Limits |
Source - originally from potable supply, then pre-treate |
Preferably reverse osmosis (RO) or filtered service deionisation (SDI) or distilled. Note: mixed bed or twin bed deionised supplies should be cation limited at exhaustion |
Fouling Index (max.) |
1 A 0.2 micron membrane prefilter is recommended for all non-RO feeds |
Service Deionization (SDI) - MO-cm |
1 MOhm-cm minimum resistivity at exhaustion |
Reverse Osmosis (RO) - µS/cm |
Recommended <30 µS/cm |
Free chlorine |
0,05 ppm max. |
TOC |
50 ppb max. |
Carbon dioxide |
30 ppm max. |
Silica |
2 ppm max. |
Particulates |
Filtration down to 0.2 micron advisable to protect internal and/or point of use filters |
temperature |
1 - 40°C, Recommended 10 - 15°C |
Flowrate (maximum requirement) |
130 l/h |
Drain requirements (gravity fall with air gap). Maximum during service |
up to 2 l/min |
Feedwater Pressure |
0,7 bar (10 psi) Maximum, 0,07 bar (1 psi) Minimum |
Electrical Requirements |
|
Mains input |
100 - 240 V AC, 50 - 60 Hz |
System voltage |
24 V DC |
Power consumption during recirculation |
60 VA |
Power consumption during dispense |
75 VA |
Fuses |
2 x T 6,3A |
Reservoir level connection |
Jack Plug 3,5 mm |
Noise level during recirculation |
<40 dBA |
1) Recommended feeding depending on the input water < 50 ppb TOC.
2) with final filter