BINDER CO2 incubator CB 220, O2 control, gas-tight, divided inner glass door

Item number: BIND_6026!

EAN: 4064343306577

Category: Incubators, Drying Cabinets

Manufacturer: Binder

Now only 15.747,11 €

Manufacturer’s suggested retail price: 16.575,90 €

Excluding 19% VAT. , plus shipping (Sperrgut)

(including 19% VAT. 18.739,05 €)
Power supply

There are variations of this item. Please select the variation you wish.


BINDER CO? incubator CB 220, O? control, gas-tight, split glass panel

The BINDER CO? incubator CB 220 with O? control and gas-tight, split glass panel was specially developed for the high-precision requirements of research laboratories, biomedical facilities, and pharmaceutical companies. The outstanding control of temperature, CO? and O? concentrations, and humidity makes this unit the preferred choice for applications that require maximum reliability and reproducibility. Compared to other CO2 incubators, the CB 220 offers an exceptional combination of intelligent design, state-of-the-art sensor technology and convenient usability - ideal for highly sensitive cell and tissue cultures.

Precision and technology for demanding applications

With an internal volume of 210 litres and sophisticated control technology, the CO2 incubator sets new standards in the field of incubation environments. The combination of a drift-free infrared CO? measuring system, optional precise oxygen control (0.2-95 vol.% O?) and effective humidification guarantees optimal conditions for cell culture and other demanding biological processes. The gas-tight, split glass panel allows users to check the culture without exposing the entire chamber to ambient conditions - a decisive advantage in avoiding fluctuations and contamination risks.

Key technical features at a glance

  • Internal volume: 210 litres
  • Temperature range: 7 °C above room temperature to 60 °C; precise, with spatial ±0.4 K and temporal ±0.1 K temperature deviation at 37 °C
  • CO2 range: 0-20 vol.%, O? range: 0.2-95 vol.%, 0.1 vol.% setting accuracy each
  • Humidity range: 90-95% RH, controlled via the Permadry dual-tank system
  • CO? and O? measurement: IR measuring system (CO2), ZrO2 technology (O2); fast regeneration times after door opening
  • Hot air sterilisation up to 180 °C, compliant with DIN 58947
  • Automatic fault diagnosis system and temperature alarm (visual and audible)

The robust, high-quality stainless steel inner chamber (560 × 750 × 500 mm) offers space for up to 8 shelves (10 kg load each, max. 30 kg total). The ergonomic design - including intuitive controls with data logger function and Ethernet connection (APT-COM DataControlSystem compatible) - enables operation and monitoring at the highest level. Intelligent features for maximum safety and convenience

The CB 220 sets new standards in terms of contamination protection and flexibility. The ANTI.PLENUM design and the VENTAIR air jacket system (depending on model) reduce potentially contamination-critical surfaces to a minimum and ensure even temperature distribution. The unit meets the highest safety requirements thanks to its temperature protection monitor and alarm management system.

The Permadry double-tank humidification system ensures defined condensation points and stable humidity control. The graphical display of all process parameters via an electronic line recorder increases controllability and enables complete documentation.

Certifications and standards

  • Hot air sterilisation in accordance with DIN 58947
  • Quality testing with BINDER test certificate
  • Temperature data in accordance with DIN 12880
  • EC conformity and safety standards for laboratory equipment

Typical areas of application

The laboratory incubator can be used universally in the natural sciences, life sciences and health sciences. The most common, but by no means exhaustive, applications include:

  • Isolation and expansion culture of human and animal cells (e.g. stem cells, immunological cell lines)
  • Cultivation of anaerobic microorganisms and sensitive bacterial strains (e.g. Escherichia coli, Lactobacillus)
  • Research and development in the field of embryonic development and IVF programmes (e.g. supported by controlled O2 concentrations)
  • Breeding of insects or small vertebrates under precisely controlled conditions (including Drosophila, zebrafish embryos)
  • Cultivation and examination of plant cells or plant-based tissue cultures in the field of agricultural biotechnology

In addition, the incubator is ideal for assisted reproduction, pharmaceutical stability testing, toxicological studies, virology and vaccine development.

FAQs

How quickly does the incubator restore the set values after the door is opened?

Regeneration times are typically 5 minutes for CO2 and temperature, and 18 minutes for O2 after a 30-second door opening.

Can the CB 220 be easily integrated into LIMS or data management systems?

Yes, thanks to its Ethernet interface and the ability to work with APT-COM DataControlSystem software, the incubator can be seamlessly integrated into digital documentation and monitoring systems.

How is the risk of cross-contamination minimised?

Key features such as the gas-tight, split glass panel, the ANTI.PLENUM design and effective hot air sterilisation offer comprehensive contamination protection. Separate humidification also supports targeted humidity control.

Is precise control of both gases guaranteed even at low settings?

Yes, control ranges of 0-20% vol. and 0.2-95% vol. are offered for both CO? and O? with an adjustment accuracy of 0.1% vol.

Conclusion

The BINDER CO? incubator CB 220 with O? control and split glass panel combines state-of-the-art technology with a well-thought-out safety and comfort concept. Its stable temperature, CO? and O? control, intelligent humidification system and extensive data management options make it the first choice for demanding applications in scientific and medical laboratory environments. With comprehensive protection mechanisms, standard-compliant sterilisation and flexible usability, it meets the highest standards and offers an outstanding platform for anyone who needs reproducible and safe cultivation processes.


Load max.: 11 - 50 kg
Width outside (incubators,...): 740 mm
Width inside (incubators,...): 560 mm
Maximum temperature (heat): + 60 °C
Height outside (incubators,...): 1070 mm
Height inside (incubators,...): 750 mm
Incubator model: CO2 incubator
Interior volume: 201 - 300 l
Cabinet type: incubator
Depth outside (incubators,...): 715 mm
Depth inside (incubators,...): 500 mm
BINDER CO2 incubator CB 220
Exterior dimensions
Width (mm) 740
Height (incl. feet) (mm) 1070
Depth (plus 54 mm for door handle and connection) (mm) 715
Wall clearance side / rear (mm) 50 / 100
Number of doors (ea.) 1
Inner glass door(s) 1
Interior dimensions
Width (mm) 560
Height (mm) 750
Depth (mm) 500
Interior volume (l) 210
Stainless steel shelf (number standard/max.) 3 / 8
Dimensions of shelf, width x depth (mm) 555,5 x 444
Load per shelf (kg) 10
Permitted total load (kg) 30
Weight (empty) (kg) 121
Temperature data
Temperature range approx. 7 °C above ambient temperature to (°C) 60
Temperature variation at 37 °C (± K) 0,4
Temperature fluctuation (± K) 0,1
Recovery time after door was opened for 30 sec. at 37 °C (min.) 1 5
Humidity data
Humidity (±2 % r.h.) 90-95
CO2 data
CO2 range (Vol.-% CO2) 0-20
Setting accuracy (Vol.-% CO2) 0,1
Recovery time after door was opened for 30 sec. 1)
at 5.0 vol. % CO2 (min.)

5
CO2 measurement IR
Connection hose nozzle for CO2 DN 6 for hose with internal diameter (mm) 6
O2 data
O2 range (Vol.-% O2) 0,2 - 95
Setting accuracy (Vol.-% O2) 0,1
Recovery time after door was opened for 30 sec. 1)
at 1.0 vol. % O2 (min.)
at 5.0 vol. % O2 (min.)

18
18
O2 measurement ZrO2
Connection hose nozzle for O2 / N2 DN 6 for hose with internal diameter (mm) 6
Electrical data
Voltage (± 10%)
50/60 Hz (V)
60 Hz (V)

230
115
Nominal power (kW) 1,5
Energy consumption at 37 °C (W) 2) 120

1) up to 98 % of the set value
2) These values can be used for dimensioning air condition systems.

All technical data is specified for unloaded units with standard equipment at an ambient temperature of 22 ± 3 °C / 71.6 ±5.4 °F and a power supply voltage fluctuation of ±10. The temperature data is determined in accordance to BINDER factory standard and DIN 12880, observing the recommended wall clearances of 10 % of the height, width and depth of the inner chamber. All indications are average values, typical for units produced in series. We reserve the right to change technical specifications at any time.

Information on product safety
Manufacturer information:
Binder GmbH
Im Mittleren Ösch 5
78532 Tuttlingen, Deutschland
[email protected]
Responsible person:
Mr. Schuldt
0049 30 403 667 940
[email protected]

Accessories