BINDER Constant climate chamber KBF LQC 720UL, 200-240 V 1 ph 50/60 Hz

Item number: S-153B!

EAN: 4064343072229

Category: BINDER

Manufacturer: Binder

32.354,00 €

Manufacturer’s suggested retail price: 32.354,00 €

Excluding 19% VAT. , plus shipping (Sperrgut)

(including 19% VAT. 38.501,25 €)
Power supply


BINDER Constant climate chamber KBF LQC 720UL, 200-240 V 1 ph 50/60 Hz
Constant climate chambers with ICH-compliant light source and light dose control

The main advantage of the BINDER KBF LQC series is the patented light photometry for photostability tests. Two 3D spherical sensors control the light dosage of UV-A and visible light. Once the desired light dosage is achieved, the chamber switches off automatically.

BENEFITS
  • Homogeneous climate conditions thanks to APT.line technology
  • Automatic water and wastewater management
  • Pressure humidification with fast response times
  • Variable position and removable illumination cassettes
  • ICH-compliant illumination for long-term stable test conditions
  • Precise photometry with 3D ball-type sensors at the sample
MAIN FEATURES
  • Temperature range: 0 °C to 70 °C
  • Temperature range with light: 10 °C to 60 °C
  • Humidity range: 10 % to 80 % RH
  • 3 positionable illumination cassettes with ICH-compliant UV/Vis light source
  • Independent light-dose control of UV-A and visible light with spherical sensors
  • APT.line preheating chamber technology
  • Humidity regulation with capacitive humidity sensor and vapor humidification
  • Inner chamber made of stainless steel
  • Intuitive touchscreen controller with time-segment and real-time programming with internal data recording
  • Tight-sealing inner door made of safety glass (ESG) with special TIMELESS coating
  • 3 stainless steel racks
  • Access port with silicone plug, 30 mm, left
  • 4 stable castors, two with brakes
  • Class 3.1 independent temperature safety device (DIN 12880) with visual and audible temperature alarm
  • Computer interface: Ethernet
  • Door heating

Load max.: 101 - 150 kg
Width outside (incubators,...): 1250 mm
Width inside (incubators,...): 973 mm
Maximum temperature (heat): + 70 °C
Height outside (incubators,...): 1925 mm
Height inside (incubators,...): 1250 mm
Interior volume: 601 - 700 l
Cabinet type: constant climate chamberclimate chamber
Depth outside (incubators,...): 890 mm
Depth inside (incubators,...): 576 mm
Lowest temperature (cooling): 0 °C
Product weight: 374,00 Kg
Dimensions: × × cm
BINDER Constant climate chamber KBF LQC 720-230V1) KBF LQC 720UL-240V1)
Performance Data Temperature
Temperature range with 100% illumination [°C] 10...60 10...60
Temperature range without illumination cassettes [°C] 0...70 0...70
Max. heat compensation at 40 °C with illumination [W] 1000 1000
Performance Data Climate
Temperature range with 100% illumination [°C] 20...60 20...60
Temperature range without illumination cassettes [°C] 10...70 10...70
Temperature variation with illumination at 25 °C and 60 % RH [± K] 1.2 1.2
Temperature variation with illumination at 40 °C and 75 % RH [± K] 1.2 1.2
Temperature fluctuation with illumination at 25 °C and 60 % RH [± K] 0.2 0.2
Temperature fluctuation with illumination at 40 °C and 75 % RH [± K] 0.2 0.2
Humidity range with illumination cassettes, with illumination [% RH] 10...75 10...75
Humidity range without illumination [% RH] 10...80 10...80
Humidity fluctuation with illumination at 25 °C and 60 % RH [± % RH] 2 2
Humidity fluctuation with illumination at 40 °C and 75 % RH [± % RH] 2 2
Humidty recovery time after 30 s door open at 25 °C and 60 % RH with illumination [min] 1 1
Humidty recovery time after 30 s door open at 40 °C and 75 % RH with illumination [min] 5 5
Lightdata per Illumination Cassettes2)
ICH compliant illumination for photo stability testing [lx] 9000 9000
ICH compliant illumination for photo stability testing [W/m2] 1.5 1.5
Electrical data
Rated Voltage [V] 200...230 200...240
Power frequency [Hz] 50 50/60
Nominal power [kW] 3.5 3.5
Unit fuse [A] 16 16
Phase (Nominal voltage) [ph] 1 1
Outer dimensions not incl. fittings and connections
Width net [mm] 1250 1250
Height net [mm] 1925 1925
Depth net [mm] 890 890
Wall clearance back [mm] 100 100
Wall clearance sidewise [mm] 200 200
Doors
Inner doors 2 2
Unit doors 2 2
Internal Dimensions
Width [mm] 973 973
Height [mm] 1250 1250
Depth [mm] 576 576
Measures
Interior volume [L] 700 700
Net weight of the unit (empty) [kg] 374 374
permitted load [kg] 150 150
Load per rack [kg] 45 45
Environment-specific data
Energy consumption at 40 °C and 75 % RH [Wh/h] 2350 2350
Sound-pressure level [dB(A)] 53 53
Fixtures
Number of shelves (std./max.) 3/12 3/12
Number of illumination cassettes (std./max.) 3/3 3/3

1) All technical data is specified for unloaded units with standard equipment at an ambient temperature of +22 °C ±3 °C and a power supply voltage fluctuation of ±10 %. The temperature data is determined in accordance to BINDER factory standard following DIN 12880, observing the recommended wall clearances of 10 % of the height, width, and depth of the inner chamber. Technical data refers to 100 % fan speed. All indications are average values, typical for units produced in series. We reserve the right to change technical specifications at any time.
2) Average value, measured at 25 °C with a spherical sensor (±10 %) by 12 cm below the light cassette. The values given in W/m2 refer to global radiation.


Accessories

Data Logger Kit TH 70 Details
Price on application
Data Logger Kit TH 70/70 Details
Price on application
Rack, stainless steel Details
Price on application
Keyboard lock Details
Price on application

Related products