BINDER Constant climate chamber KBF P 240
Item number: S-150
EAN: 4064343073745
Manufacturer:
Binder
21.306,00 €
Manufacturers suggested retail price:
21.306,00 €
Excluding 19% VAT. ,
plus
(Sperrgut)
(including 19% VAT. 25.354,13 €)
BINDER Constant climate chamber KBF P 240. Order at ProfiLab24 Online catalog for laboratory equipment. Top service. Secure payment.
- Power supply
-
There are variations of this item. Please select the variation you wish.
BINDER Constant climate chamber KBF P 240
Constant climate chambers with ICH-compliant light source
The KBF P constant climate chamber, equipped with ICH-compliant light source, is an expert when it comes to photostability tests and ensures clear test results in accordance with ICH Guideline Q1B. At the same time, the variable-position illumination cassettes offer homogeneous lighting conditions.
BENEFITS
- Homogeneous climate conditions thanks to APT.line technology
- Automatic water and wastewater management
- Pressure humidification with fast response times
- Variable position and removable illumination cassettes
- ICH-compliant illumination for long-term stable test conditions
- Precise photometry with 3D ball-type sensors at the sample
MAIN FEATURES
- Temperature range: 0 °C to 70 °C
- Temperature range with light: 10 °C to 60 °C
- Humidity range: 10 % to 80 % RH
- 2 positionable illumination cassettes with ICH-compliant UV/Vis light source
- APT.line preheating chamber technology
- Humidity regulation with capacitive humidity sensor and vapor humidification
- Inner chamber made of stainless steel
- Intuitive touchscreen controller with time-segment and real-time programming with internal data recording
- Tight-sealing inner door made of safety glass (ESG) with special TIMELESS coating
- 2 stainless steel racks
- Access port with silicone plug, 30 mm, left
- 4 stable castors, two with brakes
- Class 3.1 independent temperature safety device (DIN 12880) with visual and audible temperature alarm
- Computer interface: Ethernet
- Door heating
Load max.:
|
51 - 100 kg
|
Width outside (incubators,...):
|
925 mm
|
Width inside (incubators,...):
|
650 mm
|
Maximum temperature (heat):
|
+ 70 °C
|
Height outside (incubators,...):
|
1460 mm
|
Height inside (incubators,...):
|
785 mm
|
Interior volume:
|
201 - 300 l
|
Cabinet type:
|
constant climate chamberclimate chamber
|
Depth outside (incubators,...):
|
800 mm
|
Depth inside (incubators,...):
|
485 mm
|
Lowest temperature (cooling):
|
0 °C
|
BINDER Constant climate chamber |
KBF P 240-230V1) |
KBF P 240UL-240V1) |
Performance Data Temperature |
|
|
Temperature range with 100% illumination [°C] |
10...60 |
10...60 |
Temperature range without illumination cassettes [°C] |
0...70 |
0...70 |
Max. heat compensation at 40 °C [W] |
400 |
400 |
Performance Data Climate |
|
|
Temperature range with 100% illumination [°C] |
10...60 |
10...60 |
Temperature range without illumination cassettes [°C] |
10...70 |
10...70 |
Temperature variation with illumination at 25 °C and 60 % RH [± K] |
0.6 |
0.6 |
Temperature variation with illumination at 40 °C and 75 % RH [± K] |
0.6 |
0.6 |
Temperature fluctuation with illumination at 25 °C and 60 % RH [± K] |
0.2 |
0.2 |
Temperature fluctuation with illumination at 40 °C and 75 % RH [± K] |
0.2 |
0.2 |
Humidity range with illumination cassettes, with illumination [% RH] |
10...75 |
10...75 |
Humidity range without illumination [% RH] |
10...80 |
10...80 |
Humidity fluctuation with illumination at 25 °C and 60 % RH [± % RH] |
1.5 |
1.5 |
Humidity fluctuation with illumination at 40 °C and 75 % RH [± % RH] |
2 |
2 |
Humidty recovery time after 30 s door open at 25 °C and 60 % RH with illumination [min] |
4 |
4 |
Humidty recovery time after 30 s door open at 40 °C and 75 % RH with illumination [min] |
6 |
6 |
Lightdata per Illumination Cassettes2) |
|
|
ICH compliant illumination for photo stability testing [lx] |
7500 |
7500 |
ICH compliant illumination for photo stability testing [W/m2] |
1.1 |
1.1 |
Electrical data |
|
|
Rated Voltage [V] |
200...230 |
200...240 |
Power frequency [Hz] |
50/60 |
50/60 |
Nominal power [kW] |
2.4 |
2.4 |
Unit fuse [A] |
16 |
16 |
Phase (Nominal voltage) [ph] |
1 |
1 |
Outer dimensions not incl. fittings and connections |
|
|
Width net [mm] |
925 |
925 |
Height net [mm] |
1460 |
1460 |
Depth net [mm] |
800 |
800 |
Wall clearance back [mm] |
100 |
100 |
Wall clearance sidewise [mm] |
100 |
100 |
Doors |
|
|
Inner doors |
1 |
1 |
Unit doors |
1 |
1 |
Internal Dimensions |
|
|
Width [mm] |
650 |
650 |
Height [mm] |
785 |
785 |
Depth [mm] |
485 |
485 |
Measures |
|
|
Interior volume [L] |
247 |
247 |
Net weight of the unit (empty) [kg] |
213 |
213 |
permitted load [kg] |
100 |
100 |
Load per rack [kg] |
30 |
30 |
Environment-specific data |
|
|
Energy consumption at 40 °C and 75 % RH [Wh/h] |
760 |
760 |
Sound-pressure level [dB(A)] |
53 |
53 |
Fixtures |
|
|
Number of shelves (std./max.) |
2/7 |
2/7 |
Number of illumination cassettes (std./max.) |
2/2 |
2/2 |
1) All technical data is specified for unloaded units with standard equipment at an ambient temperature of +22 °C ±3 °C and a power supply voltage fluctuation of ±10 %. The temperature data is determined in accordance to BINDER factory standard following DIN 12880, observing the recommended wall clearances of 10 % of the height, width, and depth of the inner chamber. Technical data refers to 100 % fan speed. All indications are average values, typical for units produced in series. We reserve the right to change technical specifications at any time.
2) Average value, measured at 25 °C with a spherical sensor (±10 %) by 12 cm below the light cassette. The values given in W/m2 refer to global radiation.