Model FED056 - Drying and heating chambers Avantgarde.Line with forced convection and enhanced timer functions.
With forced convection, all BINDER drying and heating chambers Avantgarde.Line offer first-rate quality and process reliability. With its wide temperature range up to 300 °C, the drying oven can be used in all applications. Short heat up times and large power reserves are ideal for applications in science, research and industry. The excellent thermal insulation also reduces operating costs.

BENEFITS

  • Uniform drying conditions thanks to APT.lineTM technology
  • Identical test conditions throughout the chamber interior independent of sample size and quantity
  • Outstanding thermal insulation saves operation costs

Important characteristics:

  • Temperature range: ambient temperature plus 10 °C to 300 °C
  • High temperature accuracy thanks to APT.lineTM technology
  • Adjustable fan speed
  • Controller with LCD display
  • Electromechanical control of the exhaust air flap
  • 2 chrome-plated racks
  • Stackable
  • Class 2 independent temperature safety device (DIN 12880) with visual temperature alarm
  • Ergonomic handle design
  • Ethernet interface
  • USB port for recording data

Load max.: 11 - 50 kg
Width outside (incubators,...): 560 mm
Width inside (incubators,...): 400 mm
Maximum temperature (heat): + 300 °C
Height outside (incubators,...): 625 mm
Height inside (incubators,...): 440 mm
Interior volume: 51 - 100 l
Cabinet type: heating oven
Depth outside (incubators,...): 565 mm
Depth inside (incubators,...): 345 mm
Shipping weight: 60,00 Kg
BINDER Drying and heating chambers
with forced convection
FED056-230V FED056UL-120V
Option model Standard Standard
Measures - Outer dimensions
Width net (mm) 560 560
Height net (mm) 625 625
Depth net (mm) 565 565
Wall clearance back (mm) 160 160
Wall clearance sidewise (mm) 100 100
Unit doors (Stück) 1 1
Measures - Internal Dimensions
Width (mm) 400 400
Height (mm) 440 440
Depth (mm) 345 345
Interior volume (l) 60 60
Load per rack (kg) 15 15
Net weight of the unit (empty) (kg) 40 40
Performance Data Temperature
Temperature range 10 °C
above ambient temperature to (°C)
300 300
Temperature uniformity at 150 °C (± K) 1.4 1.4
Temperature fluctuation at 150 °C (± K) 0.3 0.3
Heating-up time to 150 °C (min) 15 15
Recovery time after 30 seconds
door open at 150 °C (min)
5 5
Electrical data
Rated Voltage (V) 230 120
Power frequency (Hz) 50 / 60 60
Energy consumption at 150 °C (W) 290 290
Nominal power (kW) 1.1 1.1
Unit fuse (A) 6.3 12.5
Phase (Nominal voltage) 1~ 1~