Carbolite High Temperature Clean Room Drying Oven HTCR up to 600 °C

Item number: CAR_6009

EAN: 4064343057660

Category: Carbolite

Manufacturer: Carbolite

Price on application
Model

There are variations of this item. Please select the variation you wish.


The all-rounder in research and industry: the HTCR from Carbolite

The HTCR from Carbolite is one of the most popular devices of our customers in the laboratory sector. The HTCR is the practical companion in everyday laboratory work. Your laboratory team can use it to perform a wide range of tasks. In laboratory practice or medicine, these include, for example, the cultivation of microorganisms, bacterial or animal cultures or numerous incubation or development processes. You can also use the HTCR to look at the behaviour of cells or their growth and draw specific conclusions.

The HTCR from Carbolite: the tailor-made solution for your laboratory

Incubators and incubators from the well-known manufacturer Carbolite keep their promises! The areas of application include, in particular, precise heating, drying or termperatures. And depending on the design, up to 700 °C.

Carbolite incubators comply with the BS EN 61010-2-010:2003 standard and offer excellent temperature homogeneity. It doesn't matter whether your chosen incubator is a natural or fan assisted convection unit or an atmospherically controlled high temperature oven.

Due to Carbolite's large product range, you also have many options for customisation with the HTCR. The range is almost unlimited and includes extras such as:

  • over-temperature protection
  • durable shelves and pull-outs
  • digital timers
  • ultra-modern temperature controllers

Further technical information about our HTCR

Cabinet type: drying cabinet, warming cabinetSuitable for: Drying/moisture extraction, clean room applicationsMax. load: 11 - 50 kgInternal volume: up to 50 l, 51 - 100 l, 201 - 300 l, 501 - 600 l, 901 - 1000 lAir flow: forcedWidth (outside): 810 mm, 1155 mm, 1030 mm, 1115 mm, 1530 mm, 675 mmDepth (outside): 885 mm, 1280 mm, 1120 mm, 1450 mm, 1635 mmHeight (outside): 1010 mm, 1160 mm, 880 mm, 1305 mm, 1310 mmWidth (inside): 1000 mm, 610 mm, 800 mm, 455 mm, 305 mmDepth (inside): 610 mm, 800 mm, 305 mm, 455 mm, 1000 mmHeight (inside): 800 mm, 455 mm, 610 mm, 1000 mm, 305 mmMaximum temperature (heat): + 400 °C, + 500 °C, + 600 °C

Our trained service team will be happy to provide you with more precise information about the equipment features.

Get to know even more features of the HTCR model!

Our expert team of service personnel will ensure that you receive the optimum equipment for your requirements. Do you have any questions about the versatile units, their functions or the respective brand manufacturers?

We will be happy to advise you and are always available by phone at +49 (0)30-403 667 940 or by e-mail at [email protected]. We look forward to hearing from you!


Load max.: 11 - 50 kg
Width outside (incubators,...): 675 mm810 mm1030 mm1115 mm1155 mm1530 mm
Width inside (incubators,...): 305 mm455 mm610 mm800 mm1000 mm
Maximum temperature (heat): + 400 °C+ 500 °C+ 600 °C
Height outside (incubators,...): 880 mm1010 mm1160 mm1305 mm1310 mm
Height inside (incubators,...): 305 mm455 mm610 mm800 mm1000 mm
Interior volume: up to 50 l51 - 100 l201 - 300 l501 - 600 l901 - 1000 l
Maximum power: 1 kW2 kW3 kW4 kW4.5 kW6 kW7.5 kW9 kW12 kW15 kW
Air convestion (incubators, ...): forced
Suitable for: Clean room applicationsDrying / moisture extraction
Cabinet type: heating ovenHeating cabinet
Safety standard: ISO 14644-1 Class 6
Depth outside (incubators,...): 885 mm1120 mm1280 mm1450 mm1635 mm
Depth inside (incubators,...): 305 mm455 mm610 mm800 mm1000 mm
Carbolite
Cleanroom-
Drying Cabinet
HTCR
Maximum
temperature
(°C)
Temperature-
Stability
(°C)
Temperature-
Homogeneity
(°C)
Heating time
(minutes)
Recovery-
Time
(minutes)
Dimensions
Inside
H x W x D
(mm)
Dimensions
Outside
H x W x D
(mm)
Perforated sheets
inclusive/
maximum
Maximum load
per perforated plate/
total
(kg)
Chamber-
Volume
(litres)
Maximum
power
(W)
HTCR 4/28 400 ±0,5 ±5.0 at 250 °C 60 10 305 x 305 x 305 880 x 675 x 885
(table-top unit*)
2/2 10/20 28 1000
HTCR 4/95 400 ±0,5 ±5.0 at 250 °C 90 10 455 x 455 x 455 1010 x 810 x 1120
(table-top unit*)
3/4 15/30 94 3000
HTCR 4/220 400 ±0,5 ±5.0 at 250 °C 75 16 610 x 610 x 610 1160 x 1030 x 1280
(tabletop unit*)
3/4 10/50 227 4000
HTCR 4/500 400 ±0,5 ±5.0 at 250 °C - - 800 x 800 x 800 1305 x 1115 x 1450
(floor standing unit*)
3/5 - 510 7500
HTCR 4/1000 400 ±0,5 ±5.0 at 250 °C - - 1000 x 1000 x 1000 1310 x 1530 x 1635
(floor standing unit*)
3/5 - 1000 12000
HTCR 5/28 500 ±0,5 ±5.0 at 250 °C 75 16 305 x 305 x 305 880 x 675 x 885
(table-top unit*)
2/2 10/20 28 2000
HTCR 5/95 500 ±0,5 ±5.0 at 250 °C 110 16 455 x 455 x 455 1010 x 810 x 1120
(table-top unit*)
3/4 15/30 94 4500
HTCR 5/220 500 ±0,5 ±5.0 at 250 °C 105 16 610 x 610 x 610 1160 x 1030 x 1280
(tabletop unit*)
3/4 10/50 227 6000
HTCR 5/500 500 ±0,5 ±5.0 at 250 °C - - 800 x 800 x 800 1305 x 1155 x 1450
(floor standing unit*)
3/5 10/20 510 9000
HTCR 5/1000 500 ±0,5 ±5.0 at 250 °C - - 1000 x 1000 x 1000 1310 x 1530 x 1635
(floor standing unit*)
3/5 15/30 1000 15000
HTCR 6/28 600 ±0,5 ±5.0 at 250 °C 110 20 305 x 305 x 305 880 x 675 x 885
(table-top unit*)
2/2 10/50 28 2000
HTCR 6/95 600 ±0,5 ±5.0 at 250 °C 110 20 455 x 455 x 455 1010 x 810 x 1120
(tabletop unit*)
3/4 10/20 94 4500
HTCR 6/220 600 ±0,5 ±5.0 at 250 °C 120 20 610 x 610 x 610 1160 x 1030 x 1280
(tabletop unit*)
3/4 15/30 227 9000
HTCR 6/500 600 ±0,5 ±5.0 at 250 °C - - 800 x 800 x 800 1305 x 1155 x 1450
(floor standing unit*)
3/5 - 510 12000
HTCR 6/1000 600 ±0,5 ±5.0 at 250 °C - - 1000 x 1000 x 1000 1310 x 1530 x 1635
(floor standing unit*)
3/5 - 1000 15000

Please note:
The minimum working temperature is about 60 °C above room temperature.
Temperature homogeneity measured in empty and closed chamber after a stabilisation phase
Maximum output and heating time with 240 V connection
Outer dimensions with closed door
The zone of homogeneous temperature is smaller than the total volume
* Optional with base

Power supply Volt Phase Amps per phase
HTCR 5/220 220-240 1 phase 19,5
380-415 2 phases + N 10
HTCR 6/220 220-240 1 phase 26
220-240 3 phase delta 15
380-415 3 phases + N 12,5
HTCR 6/95 220-240 1 phase 19,5
380-415 2 phases + N 10

All models not listed here in the table can be operated with a 220-240 V, 16 A single-phase connection.
Please specify desired power supply and mains connection when ordering!
Other power connection configurations on request.