RUMED explosion-proof cold/heat cabinet T 820

Item number: RUMED_34009C

EAN: 4064343052801

Category: RUMED

Manufacturer: RUMED

Price on application
Type

Product price on request.


RUMED explosion-proof cold/heat cabinet T 820
Explosion protection inside

The RUMED Safety T-Line offers a variable program of cold/heat and environmental test cabinets with explosion-proof test chambers for zone 1. Choose from 5 sizes and a multitude of options.

The units of the Safety T-Line are suitable for safe storage of explosive materials, such as aerosols, or for tests with adhesives or paint - even if the storage tests might occasionally or temporary cause an explosive atmosphere. Development of a permanent or long-term explosive atmosphere must be avoided by means of the option Technical Ventilation. The units are suitable for storage of material of the temperature classes T1, T2 and T3 of explosion group IIA, IIB and IIC. Each individual RUMED unit of the Safety T-Line is subject to an ATEX verification and receives an ATEX approval for zone 1.

Optionally available with extension of the temperature range to -20 °C/+80 °C, humidification and dehumidification and further options. Configure your test equipment, which is the most suitable for your requirements.

General view of the most important advantages:
  • Test room Ex II 2G EEx IIC T3
  • 5 Test room sizes: 320 l to 1.640 l
  • Test room and external housing of solvent-resistant stainless steel
  • Recirculating air fan for good temperature distribution in space
  • Intuitive handling by means of the Control2015 touch
  • Temperature ranges from -20 °C to +80 °C
  • Humidification and dehumidification possible
  • Possibility of calibration and validation
  • Robust and extremely durable

Width outside (incubators,...): 760 mm
Width inside (incubators,...): 610 mm
Maximum temperature (heat): + 35 °C
Height outside (incubators,...): 2105 mm
Height inside (incubators,...): 1500 mm
Interior volume: 801 - 900 l
Cabinet type: constant climate chamberclimate chambertemperature conditioning cabinetdynamic climate chamber
Depth outside (incubators,...): 1250 mm
Depth inside (incubators,...): 935 mm
Lowest temperature (cooling): 0 °C
Dimensions: × × cm
RUMED expl.-proof cold/heat cabinet Safety T-Line T 320 T 500 T 820 T 1000 T 1640
Test room explosion-proof Ex II 2 G EEx IIC T3
Volume (l) 320 500 820 1000 1650
Minimum temperature ( °C) 0 0 0 0 0
Maximum temperature ( °C) +35 +35 +35 +35 +35
Temperature deviation in time ( °C) ±0,5 ±0,5 ±0,5 ±0,5 ±0,5
Height (mm) 990 1500 1500 1500 1500
Width (mm) 610 610 610 2x 610 2x 610
Depth (mm) 585 585 935 585 935
Number of shelves (standard delivery scope) 3 4 4 8 8
Maximum load per shelf (kg) 25 25 25 25 25
Unit
Height (mm) 1600 2105 2105 2105 2105
Width (mm) 760 760 760 1520 1520
Depth (mm) 900 900 1250 900 1250
Electric connection (V/Hz) 230/50 230/50 230/50 230/50 230/50
Options
Additional shelf * * * * *
Additional shelf, reinforced (maximum load per shelf 50 kg) * * * * *
Glazed panel door, no pane heating * * * * *
Extension of temperature range down to -20 °C * * * * *
Extension of temperature range up to +80 °C 1) * * * *
Speed rate in temperature change 1 °C/min (0 °C to +35 °C) * * * * *
Speed rate in temperature change 1 °C/min (-20 °C to +80 °C) * * * * *
Ultrasonic humidification + dehumidification 2) * * * * *

*  available at extra charge
1) Electric connection is then 400 V/230 V/50 Hz, unit depth is then +225 mm
2) Unit depth is then +160 mm, removable for transportation

Katalog RUMED