Carbolite Vertical High Temperature Tube Furnace HTRV up to 1800 °C

Item number: CAR_8036

EAN: 4064343421997

Category: Carbolite

Manufacturer: Carbolite

Price on application
Selection 1-zone tube furnaces

There are variations of this item. Please select the variation you wish.


You can use our industrial furnaces for the production of different materials but also for material testing in dental laboratories.

Laboratory furnace for routine use in the laboratory

Carbolite's laboratory and industrial furnaces perform accurate work up to a good 3000 °C. In addition to reactive atmosphere, the company focuses on vacuum and inert gas. With these focal points, the devices such as the HTRV can be used especially in areas of production, research and pilot plant.

The various furnace models from Carbolite are suitable for numerous heating processes. Accordingly, the HTRV can also be used, for example, for graphitisation, drying, hardening, ashing, silicizing, sintering as well as soldering and brazing under vacuum. Do you need more detailed information about the heating processes of this model? Then please feel free to contact us.

All our Carbolite furnaces, such as the HTRV, meet excellent standards and all the necessary safety factors. In addition, you have the choice to enlarge the HTRV of the numerous digital controllers to make the tasks even more efficient. The company offers a wide range of innovative accessories for which we will be happy to advise you.

Technical information on the HTRV

Furnace unit type: Tube furnace, high-temperature furnaceSuitable for the following applications: Sintering, material testsWidth (outside): 580 mm, 455 mmHeight (outside): 515 mm, 765 mm, 365 mmDepth (outside): 455 mm, 580 mmTube length: 355 mm, 505 mm, 755 mm, 890 mm, 1040 mm, 1290 mmOven temperature max.: 1600°C, 1700°C, 1800°CPower: 12 kW, 8 kW, 3 kW, 7 kW, 10 kW, 6 kW, 2 kW, 10.4 kW, 14 kW, 4.8 kWConstant zone: 250 mm, 50 mm, 125 mmHeated length: 250 mm, 500 mm, 100 mm

For more information about our instruments, the different versions or the manufacturer, please contact our customer service!

Learn even more about the features of this version!

Our expert team of service staff will ensure that you receive the optimum equipment for your field. Do you have any questions about the various laboratory furnaces, their features or their manufacturers?

We look forward to your enquiry and will be happy to advise you. You can reach us by phone at +49 (0)30-403 667 940 or by e-mail at [email protected].


Heated length: 100 mm250 mm500 mm
Width outside (oven): 455 mm580 mm
Height outside (oven): 365 mm515 mm765 mm
Constant zone: 50 mm125 mm250 mm
Power (oven): 2 kW3 kW4.8 kW6 kW7 kW8 kW10 kW10.4 kW12 kW14 kW
Stove appliance type: high temperature furnacetube furnace
Oven temperature max.: 1600°C1700°C1800°C
Suitable for application: SinteringMaterial tests
Pipe diameter (outside): 40 mm70 mm100 mm150 mm200 mm
Pipe length: 355 mm505 mm755 mm890 mm1040 mm1290 mm
Depth outside (oven): 455 mm580 mm
Carbolite
Vertical
High temperature
Tube Furnace HTRV
Maximum
temperature

(°C)
Maximum
Outer Diameter
Working tube
(mm)
Heated
length
(mm)
Recommended
tube length under
air atmosphere
(mm)
Recommended
tube length under
inert gas atmosphere
(mm)
Dimensions
outside
H x W x D
(mm)
Weight
Oven
(kg)
Dimensions
Control unit
H x W x D
(mm)
Control unit
Weight
(kg)
Length
homogeneous zone
±5 °C
(mm)
Maximum
power
(W)
HTRV _/40/100 1600, 1700 40 100 355 890 365 x 455 x 455 30 500 x 550 x 700 50 50 2000
HTRV _/40/250 1600, 1700, 1800 40 250 505 1040 515 x 455 x 455 40 500 x 550 x 700 50 125 3000
HTRV _/40/500 1600, 1700 40 500 755 1290 765 x 455 x 455 65 850 x 550 x 700 60 250 6000
HTRV _/70/100 1600, 1700 70 100 355 890 365 x 455 x 455 30 500 x 550 x 700 50 50 3000
HTRV _/70/250 1600, 1700, 1800 70 250 505 1040 515 x 455 x 455 40 850 x 550 x 700 60 125 4800
HTRV _/70/500 1600, 1700, 1800 70 500 755 1290 765 x 455 x 455 65 850 x 550 x 700 90 250 8000
HTRV _/100/250 1600, 1700, 1800 100 250 505 1040 515 x 455 x 455 45 850 x 550 x 700 60 125 7000
HTRV _/100/500 1600, 1700, 1800 100 500 755 1290 765 x 455 x 455 70 850 x 550 x 700 90 250 10400
HTRV _/150/250 1600, 1700, 1800 150 250 505 Not available 515 x 580 x 580 55 850 x 550 x 700 90 8000
HTRV _/150/500 1600, 1700, 1800 150 500 755 Not available 765 x 580 x 580 80 850 x 550 x 700 90 12000
HTRV _/200/250 1600, 1700, 1800 200 250 505 Not available 515 x 580 x 580 70 850 x 550 x 700 90 10000
HTRV _/200/500 1600, 1700, 1800 200 500 755 Not available 365 x 580 x 580 95 850 x 550 x 700 90 14000

Please note:
Heating rate when using a ceramic working tube must not exceed +5 °C/min.
Power connection is based on 200-240 V for 1 phase and 380-415 V for 3 phases
Minimum length of homogeneous zone, with fibre plug at 100 °C below maximum temperature
Maximum continuous operating temperature is 100 °C below the maximum temperature
* In addition to the depth of the control unit, 150 mm must be added for plugs and cables.

Power supply Volt Phase Amps per phase
HTRV _/40/500 380-415 3 phases + N 16
HTRV _/70/250 380-415 3 phases + N 16
HTRV _/70/500 380-415 3 phases + N 32
HTRV _/100/250 380-415 3 phases + N 16
HTRV _/100/500 380-415 3 phases + N 32
HTRV _/150/250 380-415 3 phases + N 32
HTRV _/150/500 380-415 3 phases + N 63
HTRV _/200/250 380-415 3 phases + N 32
HTRV _/200/500 380-415 3 phases + N 63

All ovens not listed here in the table can be operated on a single-phase 220-240 V (16 A) power connection.
Please specify the desired power supply and mains connection when ordering!

Other power supply configurations on request.

Note on model selection, meaning using the example HTRV 1700/100/500
Maximum temperature: 1700 °C, Maximum outer diameter of working tube: 100 mm, Heated length: 500 mm


Accessories

Optional


Related products