Dräger Alcotest 7510

Item number: DRAEG_37004

EAN: 4026056023646

Category: Dräger

Manufacturer: Dräger

Price on application


Dräger Alcotest® 7510

Ergonomic design
The Dräger Alcotest® 7510 is a compact and robust handheld instrument which, thanks to its special ergonomic design, is suitable for right or left handed users. A non-slip rubber housing offers optimal grip.

Transflective Display
The device incorporates an innovative transflective, high resolution graphic LC display. The backlit display provides clear, unabbreviated, selectable multi-language test messages ensuring legibility in bright sunlight as well as total darkness. In addition, three light emitting diodes (LEDs) support the information displayed.

Mouth alcohol detection
The Alcotest® 7510 represents the first handheld EBT capable of detecting mouth alcohol! This technological advancement is made possible through the employement of a piezo-electric activator as part of this new sampling system.

High flexibility
The Dräger Alcotest® 7510 can be used immediately in any environment. The flexible data management, the high memory capacity as well as the option of wireless data transfer to the Dräger Mobile Printer or a computer allow for easy adjustment to national and international specifications. Via the ambient pressure sensor, the device automatically compensates for any ambient barometric conditions when performing dry-gas tests.

State-of-the-art sensor technology
The latest in Draeger sensor technology, the Alcotest® 7510 ensures short test intervals and precise measurements. A built-in heating element prevents condensation and ensures quick and accurate measurements even at sub-freezing temperatures. Draeger's fuel cell is alcohol specific and assures unsurpassed reaction and recovery performance.

Hygienic and safe
The device complies with the maximum in hygiene requirements. The sterile disposable mouthpiece employes a special tab to assure a safe distance between the operators hand and the subject's mouth. For superior hygiene, the Alcotest® 7510 has a special mouthpiece ejection mechanism. The instrument's mouthpiece receiver is iluminated for easy mouthpiece placement even in full darkness.

GPS Technology
The Alcotest® 7510 offers a GPS module recognizing the exact geographical location where a breath test is performed. Uploading these coordinates to a central computer, the administrator has the ability to establish a map showing the DUI-arrest-density. Furthermore, this technology allows for precise determination of where the test was conducted for possible administrative processing of a specific DUI stop.

Shipping weight: 1,13 Kg
Dimensions: × × cm
Alcotest 7510
Principle of measurement Electrochemical DrägerSensor in 1/4" technology; alcohol - specific
Measurement range 0 to 3.0 mg/L (if measurement range limit it exceeded a message is displayed).
Sampling Standard Sampling: Automatic when the minimum volume is reached.
Passive Sampling: without mouthpiece or manual initiation of sampling is possible.
The device can detect residual mouth alcohol during the delivery of the breath sample
(if this option is activated).
Ready for use After approximatelly 6 s of switching on (temperature-dependent)
Display of measurement result After approximatelly 3 s (at 0 mg/L); after approximatelly 10 s (at 0.5 mg/L, room temperature)
Operating temperature -10 - +50°C
Display Graphic backlit LCD display (128 x 169 pixels); 35 mm x 45 mm (1.38" x 1.77")
LED 3 light emitting diodes (LEDs) in red, yellow and green to support display and warning messages
Audible Signal Different signal tones to support display messages and warnings
Datalogger Storage of > 5,000 tests with test numbers, date and time
Power supply Four AA alkaline or NiMH batteries, charge status is displayed.
Approximately 1,500 breath tests can be performed with one set of batteries.
Optional: Lithium-Ion batteries can perform approximatelly
50% more measurements than with NiMH rechargeable batteries.
Calibration Wet gas or dry gas calibration.
Optional: with absolute pressure sensor for the
compensation of atmospheric pressure during dry gas calibration
Dimensions (H x W x T), Weight Approximatelly 185 mm x 90 mm x 45 mm (7.3" x 3.5" x 1.77")
Approximatelly 430g (0.96 lbs)
Instrument configuration Direct menu-guided configuration of instrument settings (PIN required). Additional configuration via PC software
Vibration und shock Conforms to EN 60068-2-6, EN 60068-2-29
CE marking Directive 89/336/EC (electromagnetic compatibility) RoHs Complianc
Data Interfaces IR (via IR Module) and USB for connection to a PC;
IR connection to the Dräger Mobile Printer for documentation and data entry
(when a keyboard is attached to the Dräger Mobile Printer)