Socorex Acura® manual 825 adjustable volume micropipettes - ProfiLab24 GmbH
99
100
24
4
stars -
Professional service
Customer
Young team with a lot of ambition and quick answers to all questions. We needed a complete laboratory equipment for our new location. The team of Profiab24 has arranged everything.
Socorex Acura® manual 825 adjustable volume micropipettes
starting from
275,19 €
Manufacturers suggested retail price:
289,67 €
Excluding 19% VAT. ,
plus
(Standard Light)
(including 19% VAT. 327,47 €)
Socorex Acura® manual 825 adjustable volume micropipettes at ProfiLab24 dealer for laboratory equipment. Made in Germany with top advice and 24H service.
- Volume [µl]
-
There are variations of this item. Please select the variation you wish.
Acura® manual 825 adjustable volume micropipettes are equipped with a precision setting mechanism, the line is continuously adjustable down to a very fine volume resolution. Single-handed volume setting easy when other hand busy. Click stops, together with freely rotating Smartie plunger button cap secure setting stability while pipetting.
A total of 8 micropipettes covering volumes from 0.1 to 1000 µL.
Highlights
- Engineered ergonomics and design
- Smart and reliable volume setting
- Display always visible at a glance
- Tip ejection - easier than ever
- Swift-set calibration
- Colour coding at will
- Minimum maintenance / sterilisation
- QC certificate
The Acura® manual family was built with the user in mind. Instruments combine high-tech materials with exceptional user friendly ergonomics. Whether in research or routine applications they bring more than superior performance and make pipetting safe and agreeable.
- Excellence in ergonomics
- Light weight and softness in use
- Unsurpassed metrologic performance and stability
- Adjustable tip ejector (Socorex patented) fitting most tips
- Swift-set user calibration system (Socorex patented)
- Shock, UV-light and autoclaving resistance
- CE certified IVD 98/79 EEC
Pipetting volume:
|
variable
|
Type (pipettes ...):
|
micropipette
|
Shipping weight: |
1,50 Kg |
|
Inaccuracy (Rel. err.%) |
Imprecision (CV %) |
|
Vol. µl |
Div. µl |
Min. vol. |
Mid. vol. |
Max. vol. |
Min. vol. |
Mid. vol. |
Max. vol. |
Tip style |
0.1 - 2 |
0.002 |
<±6.0%* |
<± 4.0% |
<± 2.0% |
<5.0%* |
< 3.3% |
< 1.5% |
ultra 10 µL |
0.5 - 10 |
0.01 |
<±2.5%** |
<± 1.8% |
<± 1.0% |
<1.8%** |
< 1.2% |
< 0.5% |
ultra 10 µL |
1 - 10 |
0.01 |
<± 2.5% |
<± 1.8% |
<± 1.0% |
< 2.5% |
< 1.6% |
< 0.7% |
200 µL |
2 - 20 |
0.02 |
<± 2.5% |
<± 1.8% |
<± 1.0% |
< 1.7% |
< 1.1% |
< 0.5% |
200 µL |
5 - 50 |
0.1 |
<± 1.5% |
<± 1.3% |
<± 1.0% |
< 1.0% |
< 0.7% |
< 0.4% |
200 µL |
10 - 100 |
0.1 |
<± 1.5% |
<± 1.2% |
<± 0.8% |
< 1.0% |
< 0.6% |
< 0.2% |
200 µL |
20 - 200 |
0.2 |
<± 1.5% |
<± 1.1% |
<± 0.6% |
< 0.6% |
< 0.4% |
< 0.2% |
200 µL |
100 - 1000 |
1 |
<± 1.5% |
<± 1.0% |
<± 0.5% |
< 0.5% |
< 0.4% |
< 0.2% |
1000 µL |
Measured at * 0.5µL, ** 1µL
Performance values obtained with bi-dist. water at constant temperature (± 0.5°C) comprised between 20 and 25°C in accordance with ISO 8655.