ELGA PURELAB 7060

Item number: ELGA_21045>

Category:

ELGA

Manufacturer: VWS Deutschland

Price on application

Product price on request.


ELGA PURELAB 7060

The PURELAB 7000 series is designed for applications requiring higher flows of Type II pure water or as a pure water feed system for small distribution loops. The system can supply water from the internal 50 liter reservoir at varying flow rates or can feed into much larger vessels as required.
  • Small high purity distribution system with variable flow rates
  • Auto rinse and simple sanitization procedures maintain the purity of water for the life of the system


For operating a PURELAB 3000/7000 system, a starter kit 1 or 2 is required.

Modell: 7060
Water treatment model: pure water system
Booster pump: without
Capacity: 60 l/h
Storage tank - tank: with
Water source: of ion exchanger from the water mains
Product weight:60,00 Kg
Model PURELAB 7060
Treated Water Specifications
Make-up rate at 15°C60 l/hr
Daily usage - typical480 liters
Daily usage - maximum1440 liters
Delivery flowrateUp to 4 l/min at 12 psi (0.8 bar) 1)
Inorganics rejection-
Inorganics (resistivity at 25°C)>15 MOhm-cm
Bacteria - typical<1 CFU/ml
Organics (TOC) - typical<10 ppb
Particles0.05 µm
Dimensions and weights
DimensionsHeight 820mm (32.8in)
834mm (33.4in) including castors
Width 794mm (31.8in)
Depth 470mm (18.8in)
Supply weight52kg
Operational weight109kg
InstallationFloor/bench
Feedwater Requirements
SourceTap water as detailed below
Contaminant
Conductivity<2000 µS/cm*
Hardness<250 ppm as CaCO3
Free chlorine<4 ppm Cl2
Chloramine<1 ppm as Cl2
Silica<30 ppm SiO2
Fouling Index<10 FI
Iron/manganese<0.1 ppm Fe/Mn
Organics<3 ppm TOC
Temperature4 - 40°C
(Recommended 15 - 25°C)
Flowrate (maximum requirement at 15°C)9 l/min
Drain requirements (gravity fall with air gap)20 l/min
Feedwater pressure6 bar (90 psi) maximum
2 bar (30 psi) minimum
Electrical Requirements
Mains input230V ac, 50Hz
115V ac, 60Hz
System control voltage (not including pumps and UV)24V dc
Power consumption (peak demand)650VA
Electrical protection rating10 amps


Specifications are for a system fed with a suitable water supply and installed, operated and sanitized according to the operator manuals.
1) With High Flow Kit (LA723) or 2 l/min at 12 psi (0.8 bar) without high flow kit

Failure to comply with the feedwater pretreatment recommendations will affect the life and performance of key components and may invalidate the warranty. Feedwater requirements are specified for units fitted with a Protek L2 pretreatment cartridge. If feedwater purity is variable or values are outside any of the above ranges, additional pretreatment is recommended to be installed in the feedwater supply to the unit.

*
Deionization cartridge life may vary with feedwater >1400 µS/cm