Features
- Delivers up to 60 liters an hour of type II water
- Up to 15 MOhm-cm, type II water suitable for general laboratory applications
- Unique recirculating Type II system guarantees water quality at dispense
- Easy to maintain
- Full service and validation support
To operate a PURELAB Option-S / R Systems, a tank with 3 level switches is required
|
Booster pump:
|
without
|
|
Power:
|
9 l/min
|
|
Modules:
|
UV-modulesRO-module
|
|
Storage tank - tank:
|
without
|
|
Water treatment type:
|
pure water system
|
|
Water source:
|
from the water mains
|
| Shipping weight: |
54,00 Kg |
| Product weight: |
54,00 Kg |
| Model |
Option-R 60 |
| Treated Water Specifications |
|
| Make up rate 1) |
60 l/hr |
| Dispense rate from tap |
1.0 l/min-nominal (less with POU filter) |
| Purity |
|
| Inorganics @ 25°C |
5 - 15 MOhm-cm |
| Total organic carbon (TOC) |
<20 ppb |
| Bacteria 2)
|
<1 CFU/ml |
| pH |
Effectively neutral |
| Ion/organic cartridge capacity |
165,000 liters >5 MO-cm µS/cm at pH7 3)
|
| Dimensions and weights |
|
| Height |
740mm |
| Width |
570mm |
| Depth |
320mm |
| Weight (operational) |
54kg |
| Feedwater Requirements |
|
| Source quality |
Potable mains water supply |
| Fouling index - maximum |
10 |
| Total dissolved solids - maximum |
1400µS/cm |
| Free chlorine - maximum |
<0.5 ppm |
| Heavy metals - maximum |
0.05 ppm |
| Temperature |
1 - 35°C |
| Flowrate (maximum requirement) |
9 l/min |
| Feedwater pressure maximum |
6.0 bar (90 psi) |
| Feedwater pressure minimum |
1.4 bar (20 psi) |
| Electrical Requirements |
|
| Mains input |
230V ac, 50Hz 115V ac, 60Hz 100V ac, 50/60Hz |
| Power consumption with boost pump |
700VA |
| Fuses |
2 x 3.15 Amp - 230V ac, 50Hz 2 x 6.3 Amp - 115V ac, 60Hz |
| Reservoir level connection |
Din plug 6 way |
| Noise level |
<52dBA |
1) Subject to suitable feedwater
2) Subject to suitable operating procedures
3)Conductivity of feed to Deionization Cartridge (RO Quality)