ELGA water treatment system PURELAB Chorus 1 - ProfiLab24 GmbH
99
100
24
4
stars -
Professional service
Customer
Young team with a lot of ambition and quick answers to all questions. We needed a complete laboratory equipment for our new location. The team of Profiab24 has arranged everything.
ELGA water treatment system PURELAB Chorus 1
starting from
5.377,00 €
Manufacturers suggested retail price:
5.660,00 €
Excluding 19% VAT. ,
plus
(Standard)
(including 19% VAT. 6.398,63 €)
ELGA water treatment system PURELAB Chorus 1
The compact PURELAB Chorus 1 produces up to 120 litres of ultra-pure water of the highest quality (type I+) per day from pre-treated feed water for demanding applications.
The two-stage PureSure deionisation removes even residual ions for precise results.
It avoids interrupted workflows due to early notice of cartridge changes and saves costs by maximising capacity utilisation.
An ultramicrofilter removes bacteria and particles (Analytical Research), the ultrafiltration module also retains endotoxins and nucleases (Life Science).
Easy system cleaning with disinfection cartridge.
- Two-stage PureSure deionisation
- Full spectrum UV lamp
- Real-time TOC monitoring
- Full recirculation
- Data logging for validation
Width (water treatment):
|
375 mm
|
Booster pump:
|
without
|
Height (water treatment):
|
435 mm
|
Ideal for:
|
ICP and ICP/MSHPLCMolecular biologyUltra-trace analysisCell cultures
|
Power:
|
2 l/min
|
Modules:
|
UV lamp 185/254 nmultrafiltration module
|
Depth (water treatment):
|
340 mm
|
Storage tank - tank:
|
without
|
Water treatment type:
|
ultrapure water system
|
Water source:
|
from the water mains
|
Dimensions: |
× × cm |
ELGA PURELAB |
Chorus 1 General Science |
Chorus 1 Analytical Research |
Chorus 1 Life Science |
Withdrawal flow rate |
up to 2.0 l/min (with Halo Dispenser) |
up to 2.0 l/min (with Halo Dispenser) |
up to 2.0 l/min (with Halo Dispenser) |
Inorganic substances at 25 °C |
18.2 mega-ohm x cm, 0.055 µS/cm |
18.2 mega-ohm x cm, 0.055 µS/cm |
18.2 mega-ohm x cm, 0.055 µS/cm |
Organic substances (TOC) |
3-10 ppb (depending on feed water) |
1-3 ppb (depending on feed water) |
1-3 ppb (depending on feed water) |
Bacteria |
<0.001 CFU/ml
(with POU filter LC134/LC145/LC197) |
<0.001 CFU/ml
(with POU filter LC134/LC145/LC197) |
<0.001 CFU/ml
(with POU filter LC134/LC145/LC197) |
Bacterial endotoxins |
<0.001 EU/ml (with POU filter LC197) |
<0.001 EU/ml (with POU filter LC197) |
<0.001 EU/ml |
Particle removal |
0.2 µm
(with POU filter LC134/LC145/LC197) |
0.05 µm |
<0.01 µm |
Rnase |
|
<1 pg/ml |
<1 pg/ml |
Dnase |
|
<5 pg/ml |
<5 pg/ml |
Fouling index |
1 maximum |
1 maximum |
1 maximum |
Ion exchanger SDI |
1 mega-ohm x cm minimum resistance at exhaustion |
1 mega-ohm x cm minimum resistance at depletion |
1 mega-ohm x cm minimum resistance at exhaustion |
Reverse osmosis (RO) |
recommended <30 µS/cm |
recommended <30 µS/cm |
recommended <30 µS/cm |
Free chlorine |
0.05 ppm maximum |
0.05 ppm maximum |
0.05 ppm maximum |
TOC |
50 ppb maximum (reverse osmosis feed) |
50 ppb maximum (reverse osmosis feed) |
50 ppb maximum (reverse osmosis feed) |
Carbon dioxide |
30 ppm maximum |
30 ppm maximum |
30 ppm maximum |
Silicate |
2 ppm maximum |
2 ppm maximum |
2 ppm maximum |
Particles |
a 0.2 µm membrane prefilter is recommended |
a 0.2 µm membrane prefilter is recommended |
a 0.2 µm membrane prefilter is recommended |
Temperature |
1-35 ºC (10-15 ºC recommended) |
1-35 ºC (recommended 10-15 ºC) |
1-35 ºC (recommended 10-15 ºC) |
Flow rate |
130 l/h maximum |
130 l/h maximum |
130 l/h maximum |
Feed water pressure |
minimum 0.07 bar, maximum 0.7 bar
(at >0.7 bar a pressure reducer is required,
optional accessory) |
minimum 0.07 bar, maximum 0.7 bar
(at >0.7 bar, a pressure reducer is required,
optional accessories) |
minimum 0.07 bar, maximum 0.7 bar
(a pressure reducer is required for >0.7 bar,
optional accessories) |
Outlet (free outlet) |
up to 2 l/min |
up to 2 l/min |
up to 2 l/min |
Installation |
on / under table or wall |
on / under table or wall |
on / under table or wall |
Mains voltage |
100-240 V AC, 50-60 Hz |
100-240 V AC, 50-60 Hz |
100-240 V AC, 50-60 Hz |
Power input |
maximum 90 W |
maximum 90 W |
maximum 90 W |
System voltage |
24 V DC |
24 V DC |
24 V DC |
Low-noise operation at |
<40 dBA |
<40 dBA |
<40 dBA |
|
|
|
|
Height |
435 mm |
435 mm |
435 mm |
Width |
375 mm |
375 mm |
375 mm |
Depth |
340 mm |
340 mm |
340 mm |
Weight |
18 kg |
19 kg |
19 kg |
PURELAB Chorus 1 systems are delivered including 2 x LC232 preparation cartridges (RO feed), 1 x LA762 installation kit Basic and 2 x LA233 bypass block.
A separately available Halo Dispenser is required to operate a PURELAB Chorus 1 system.
Up to four Halo Dispensers can be connected to one PURELAB Chorus 1 system.