ELGA water treatment system PURELAB Chorus 2 (RO/DI)

Item number: ELGA_21072

EAN: 4064343088381

Category: ELGA

Manufacturer: VWS Deutschland

starting from 4.509,65 €

Manufacturer’s suggested retail price: 4.747,00 €

Excluding 19% VAT. , plus shipping (Standard)

(including 19% VAT. 5.366,48 €)
Version
Power supply

There are variations of this item. Please select the variation you wish.


ELGA water treatment system PURELAB Chorus 2 (RO/DI)

The compact system produces type 2 pure water from untreated drinking water.
For higher water requirements, up to four PURELAB Chorus 2 RO/DI systems can be combined in a modular system.
The colour display shows tank level, product water quality and temperature, permate quality and system messages. Alarm values are freely selectable for product water quality (1, 10 or 15 mega-ohm cm) and temperature (30, 35 or 40 °C).
The system can be combined with Chorus storage tanks and networked with other Chorus treatment systems.

  • Pure water (10 mega-ohms) directly from drinking water
  • Space-saving installation options: Wall, table, under-table, on and under tanks
  • Economical thanks to optimised capacity
  • Easy to upgrade to higher flow rates

Width (water treatment): 375 mm
Booster pump: with
Height (water treatment): 435 mm
Ideal for: Feeding ultrapure water systemsMedia productionBuffer approachReagentsClimate and salt spray chambersph solutionsclean laboratory environmentas a substitute for stillsWashing / Rinsing
Power: 10 l/h20 l/h
Modules: RO-module
Depth (water treatment): 340 mm
Storage tank - tank: optional
Water treatment type: RO waterpure water systemReverse osmosis system
Water source: from reverse osmosis systemfrom the water mains
Dimensions: × × cm
ELGA PURELAB Chorus 2 (RO/DI) 10 l/h Chorus 2 (RO/DI) 10 l/h, BP Chorus 2 (RO/DI) 20 l/h Chorus 2 (RO/DI) 20 l/h, BP
Production rate maximum 10 l/h (240 l/24h)
(4 bar inlet pressure at 15 °C)
maximum 10 l/h (240 l/24h)
(4 bar inlet pressure at 15 °C)
maximum 20 l/h (480 l/24h)
(4 bar inlet pressure at 15 °C)
maximum 20 l/h (480 l/24h)
(4 bar inlet pressure at 15 °C)
Inorganic substances at 25 °C 1 to >10 mega-ohm x cm 1 to >10 mega-ohm x cm 1 to >10 mega-ohm x cm 1 to >10 mega-ohm x cm
Organic substances (>200 Da) >99% retention rate >99% retention rate >99% retention rate >99% retention rate
Organic matter (TOC) <30 ppb <30 ppb <30 ppb <30 ppb
Bacteria <5 CFU/ml <5 CFU/ml <5 CFU/ml <5 CFU/ml
pH value neutral neutral neutral neutral
Particle >99% retention rate >99% retention rate >99% retention rate >99% retention rate
Fouling Index 10 maximum 10 maximum 10 maximum 10 maximum
Conductivity <2,000 µS/cm <2,000 µS/cm <2,000 µS/cm <2,000 µS/cm
Free chlorine 0.5 ppm maximum 0.5 ppm maximum 0.5 ppm maximum 0.5 ppm maximum
Heavy metals 0.05 ppm maximum 0.05 ppm maximum 0.05 ppm maximum 0.05 ppm maximum
Silicate 30 ppm 30 ppm 30 ppm 30 ppm
Temperature 1-35 ºC 1-35 ºC 1-35 ºC 1-35 ºC
Flow rate 100 l/h maximum 100 l/h maximum 100 l/h maximum 100 l/h maximum
Feed water pressure minimum 2 bar, maximum 6.0 bar
(at >6 bar a pressure reducer is required,
optional accessory)
minimum 0.5 bar, maximum 2.0 bar
(at >2 bar a pressure reducer is required,
optional accessories)
minimum 2 bar, maximum 6.0 bar
(a pressure reducer is required for >6 bar,
optional accessories)
minimum 0.5 bar, maximum 2.0 bar
(at >2 bar, a pressure reducer is required,
optional accessories)
Outlet (free outlet) 80 l/h 80 l/h 80 l/h 80 l/h
Installation on / under table or wall on / under table or wall on / under table or wall on / under table or wall
Mains voltage 100-240 V AC, 50-60 Hz 100-240 V AC, 50-60 Hz 100-240 V AC, 50-60 Hz 100-240 V AC, 50-60 Hz
Power input maximum 50 W maximum 50 W maximum 50 W maximum 50 W
System voltage 24 V DC 24 V DC 24 V DC 24 V DC
Low-noise operation at <40 dBA <40 dBA <40 dBA <40 dBA
Height 435 mm 435 mm 435 mm 435 mm
Width 375 mm 375 mm 375 mm 375 mm
Depth 340 mm 340 mm 340 mm 340 mm
Weight 17 kg 19 kg 18 kg 20 kg


PURELAB Chorus 2 (RO/DI) systems are delivered ready for operation including LC241 pre-treatment cartridge, 1-2 x LC240 RO module, LC232 treatment cartridge, LA762 installation kit Basic and LA233 bypass block.
A PURELAB Chorus storage tank is required to operate a PURELAB Chorus 2 (RO/DI) system.