(including 19% VAT. 6.850,83 €)
- Model
-
There are variations of this item. Please select the variation you wish.
ELGA water treatment system PURELAB Chorus 2+ (RO/DI/UV).
The compact system produces type 2+ pure water from untreated drinking water at up to 20 l/h (withdrawal >1.5 l/min).
For constant high water quality with low bacteria content, the treated water recirculates from the tank to the withdrawal via the deionisation and the UV lamp.
The colour display shows water parameters, status messages and the tank level at a glance.
Alarm values can be selected for permeate quality, product water quality and temperature.
The system can be optionally combined with up to three Halo dispensers for pure water extraction and final filters.
- Pure water (15 mega-ohms) directly from drinking water
- Recirculation for guaranteed purity
- Space-saving installation options: Wall, table, under-table, on tanks
Width (water treatment):
|
376 mm
|
Booster pump:
|
withoutwith
|
Height (water treatment):
|
679 mm
|
Ideal for:
|
ElectrochemistryBuffer approachReagentsph solutionsclean laboratory environmentCell culturesSpectrophotometryWashing / Rinsingcytological applicationshistological applications
|
Power:
|
10 l/h20 l/h
|
Modules:
|
UV lamp 254 nmReverse osmosis
|
Depth (water treatment):
|
353 mm
|
Storage tank - tank:
|
optional
|
Water treatment type:
|
pure water systemfilter systemReverse osmosis system
|
Water source:
|
from reverse osmosis systemfrom the water mains
|
Dimensions: |
× × cm |
ELGA PURELAB |
Chorus 2+ (RO/DI/UV) 10 l/h |
Chorus 2+ (RO/DI/UV) 10 l/h, BP |
Chorus 2+ (RO/DI/UV) 20 l/h |
Chorus 2+ (RO/DI/UV) 20 l/h, BP |
Production rate |
maximum 10 l/h (240 l/24h)
(4 bar inlet pressure at 15 °C) |
maximum 10 l/h (240 l/24h) |
maximum 20 l/h (480 l/24h)
(4 bar inlet pressure at 15 °C) |
maximum 20 l/h (480 l/24h) |
Flow rate at the tap |
up to 1.5 l/min (without POU filter) |
up to 1.5 l/min (without POU filter) |
up to 1.5 l/min (without final filter) |
up to 1.5 l/min (without POU filter) |
Inorganic substances at 25 °C |
10 to >15 mega-ohm x cm |
10 to >15 mega-ohm x cm |
10 to >15 mega-ohm x cm |
10 to >15 mega-ohm x cm |
Organic substances (>200 Da) |
>99% retention rate |
>99% retention rate |
>99% retention rate |
>99% retention rate |
Organic matter (TOC) |
<10 ppb |
<10 ppb |
<10 ppb |
<10 ppb |
Bacteria |
<0.001 CFU/ml (with POU filter LC134/LC145) |
<0.001 CFU/ml (with POU filter LC134/LC145) |
<0.001 CFU/ml (with microfilter LC145) |
<0.001 CFU/ml (with POU filter LC134/LC145) |
PH value |
neutral |
neutral |
neutral |
neutral |
Particles |
0.2 µm (with POU filter LC134/LC145) |
0.2 µm (with POU filter LC134/LC145) |
0.2 µm (with micro filter LC145) |
0.2 µm (with POU filter LC134/LC145) |
Fouling index |
10 maximum |
10 maximum |
10 maximum |
10 maximum |
Conductivity |
<2,000 µS/cm |
<2,000 µS/cm |
<2,000 µS/cm |
<2,000 µS/cm |
Free chlorine |
0.5 ppm maximum |
0.5 ppm maximum |
0.5 ppm maximum |
0.5 ppm maximum |
Heavy metals |
0.05 ppm maximum |
0.05 ppm maximum |
0.05 ppm maximum |
0.05 ppm maximum |
CO2 |
<20 ppm (recommendation) |
<20 ppm (recommendation) |
<20 ppm (recommendation) |
<20 ppm (recommendation) |
Silicate |
30 ppm |
30 ppm |
30 ppm |
30 ppm |
Temperature |
1-35 °C |
1-35 °C |
1-35 °C |
1-35 °C |
Flow rate |
100 l/h maximum |
100 l/h maximum |
100 l/h maximum |
100 l/h maximum |
Feed water pressure |
minimum 4.0 bar, maximum 6.0 bar
(at >6 bar a pressure reducer is required,
optional accessory) |
minimum 0.5 bar, maximum 2.0 bar
(at >2 bar a pressure reducer is required,
optional accessories) |
minimum 4.0 bar, maximum 6.0 bar
(a pressure reducer is required for >6 bar,
optional accessories) |
minimum 0.5 bar, maximum 2.0 bar
(at >2 bar, a pressure reducer is required,
optional accessories) |
Outlet (free outlet) |
80 l/h |
80 l/h |
80 l/h |
80 l/h |
Installation |
on/under table or wall |
on/under table or wall |
on/under table or wall |
on/under table or wall |
Mains voltage |
100-240 V AC, 50-60 Hz |
100-240 V AC, 50-60 Hz |
100-240 V AC, 50-60 Hz |
100-240 V AC, 50-60 Hz |
Input power |
maximum 155 VA |
Maximum 155 VA |
maximum 155 VA |
maximum 155 VA |
System voltage |
24 V DC |
24 V DC |
24 V DC |
24 V DC |
Low-noise operation at |
<45 dBA |
<45 dBA |
<45 dBA |
<45 dBA |
|
|
|
|
|
Height |
679 mm |
679 mm |
679 mm |
679 mm |
Width |
376 mm |
376 mm |
376 mm |
376 mm |
Depth |
353 mm |
353 mm |
353 mm |
353 mm |
Weight |
15 kg |
17 kg |
16 kg |
18 kg |
PURELAB Chorus 2+ (RO/DI/UV) systems are delivered ready for operation including LC241 pre-treatment cartridge, 1-2 x LC240 RO module, LC274 treatment cartridge, LC285 UV lamp,
LA762 Basic Installation Kit, LA233 Bypass Block and 1 x POWE40855 Power Supply.