ELGA water treatment system PURELAB Chorus 2+ (RO/EDI/UV)
starting from
8.142,45 €
Manufacturers suggested retail price:
8.571,00 €
Excluding 19% VAT. ,
plus
(Standard)
(including 19% VAT. 9.689,52 €)
- Model
-
There are variations of this item. Please select the variation you wish.
ELGA water treatment system PURELAB Chorus 2+ (RO/EDI/UV).
The compact system produces type 2+ pure water from untreated drinking water at up to 20 l/h (withdrawal >1 l/min).
For constant water quality and low operating costs at high demand, the treated water recirculates from the tank to withdrawal via the patented electrodeionisation (EDI) and UV lamp.
The colour display shows water parameters, status messages and tank level at a glance.
Alarm values can be selected for permeate quality, product water quality and temperature.
The system can be optionally combined with up to three halo dispensers for pure water extraction as well as final filters.
- Pure water (> 15 mega-ohms) directly from drinking water
- Cost-effective for high demands thanks to EDI
- Space-saving installation options: Wall, table, under-table, on tanks
- Recirculation for guaranteed purity
Width (water treatment):
|
376 mm
|
Booster pump:
|
withoutwith
|
Height (water treatment):
|
679 mm
|
Ideal for:
|
ReagentsFeeding ultrapure water systemsph solutionsMedia productionCell culturesBuffer approachSpectrophotometryElectrochemistryWashing / Rinsingcytological applicationshistological applications
|
Power:
|
10 l/h20 l/h
|
Modules:
|
Membrane degassingEDI-moduleUV lamp 254 nmReverse osmosis
|
Depth (water treatment):
|
353 mm
|
Storage tank - tank:
|
optional
|
Water treatment type:
|
pure water systemReverse osmosis system
|
Water source:
|
from reverse osmosis systemfrom the water mains
|
Dimensions: |
× × cm |
ELGA PURELAB |
Chorus 2+ (RO/EDI/UV) 10 l/h |
PURELAB Chorus 2+ (RO/EDI/UV) 10 l/h, BP |
PURELAB Chorus 2+ (RO/EDI/UV) 20 l/h |
PURELAB Chorus 2+ (RO/EDI/UV) 20 l/h, BP |
Production rate |
maximum 10 l/h (220 l/24h)
(4 bar inlet pressure at 15 °C) |
maximum 10 l/h (220 l/24h) |
maximum 20 l/h (220 l/24h)
(4 bar inlet pressure at 15 °C) |
Maximum 20 l/h (220 l/24h) |
Flow rate at the tap |
up to 1.0 l/min (without POU filter) |
up to 1.0 l/min (without POU filter) |
up to 1.0 l/min (without POU filter) |
up to 1.0 l/min (without POU filter) |
Inorganic substances at 25 °C |
10 to >15 mega-ohm x cm |
10 to >15 mega-ohm x cm |
10 to >15 mega-ohm x cm |
10 to >15 mega-ohm x cm |
Organic substances (>200 Da) |
>99% retention rate |
>99% retention rate |
>99% retention rate |
>99% retention rate |
Organic matter (TOC) |
<10 ppb |
<10 ppb |
<10 ppb |
<10 ppb |
Bacteria |
<1 CFU/ml (without POU filter) |
<1 CFU/ml (without POU filter) |
<1 CFU/ml (without POU filter) |
<1 CFU/ml (without POU filter) |
|
<0.001 CFU/ml (with POU filter LC134/LC145) |
<0.001 CFU/ml (with POU filter LC134/LC145) |
<0.001 CFU/ml (with POU filter LC134/LC145) |
<0.001 CFU/ml (with POU filter LC134/LC145) |
PH value |
neutral |
neutral |
neutral |
neutral |
Particles |
0.2 µm (with POU filter LC134/LC145) |
0.2 µm (with POU filter LC134/LC145) |
0.2 µm (with POU filter LC134/LC145) |
0.2 µm (with POU filter LC134/LC145) |
Fouling index |
10 maximum |
10 maximum |
10 maximum |
10 maximum |
Conductivity |
<1,400 µS/cm |
<1,400 µS/cm |
<1,400 µS/cm |
<1,400 µS/cm |
Free chlorine |
0.5 ppm maximum |
0.5 ppm maximum |
0.5 ppm maximum |
0.5 ppm maximum |
Heavy metals |
0.05 ppm maximum |
0.05 ppm maximum |
0.05 ppm maximum |
0.05 ppm maximum |
CO2 |
<20 ppm (recommendation) |
<20 ppm (recommendation) |
<20 ppm (recommendation) |
<20 ppm (recommendation) |
|
<30 ppm maximum |
<30 ppm maximum |
<30 ppm maximum |
<30 ppm maximum |
Silicate |
30 ppm |
30 ppm |
30 ppm |
30 ppm |
Temperature |
1-35 °C |
1-35 °C |
1-35 °C |
1-35 °C |
Flow rate |
100 l/h maximum |
100 l/h maximum |
100 l/h maximum |
100 l/h maximum |
Feed water pressure |
minimum 4.0 bar, maximum 6.0 bar
(at >2 bar a pressure reducer is required,
optional accessory) |
minimum 0.5 bar, maximum 2.0 bar
(at >2 bar a pressure reducer is required,
optional accessories) |
minimum 4.0 bar, maximum 6.0 bar
(a pressure reducer is required for >2 bar,
optional accessories) |
minimum 0.5 bar, maximum 2.0 bar
(at >2 bar, a pressure reducer is required,
optional accessories) |
Outlet (free outlet) |
80 l/h |
80 l/h |
80 l/h |
80 l/h |
Installation |
on / under table or wall |
on / under table or wall |
on / under table or wall |
on / under table or wall |
Mains voltage |
100-240 V AC, 50-60 Hz |
100-240 V AC, 50-60 Hz |
100-240 V AC, 50-60 Hz |
100-240 V AC, 50-60 Hz |
Input power |
maximum 155 VA |
Maximum 155 VA |
maximum 155 VA |
maximum 155 VA |
System voltage |
24 V DC |
24 V DC |
24 V DC |
24 V DC |
Low-noise operation at |
<45 dBA |
<45 dBA |
<45 dBA |
<45 dBA |
|
|
|
|
|
Height |
679 mm |
679 mm |
679 mm |
679 mm |
Width |
376 mm |
376 mm |
376 mm |
376 mm |
Depth |
353 mm |
353 mm |
353 mm |
353 mm |
Weight |
18 kg |
21 kg |
19 kg |
22 kg |
PURELAB Chorus 2+ (RO/EDI/UV) systems are delivered ready for operation including LC141 pre-treatment cartridge, 1-2 x LC240 RO module, LC243 conditioning cartridge, LC277 EDI module, LC285 UV lamp,
LA762 Installation Kit Basic, LA233 Bypass Block and 1 x POWE40855 Power Supply.
A PURELAB Chorus supply tank is required to operate a PURELAB Chorus 2+ (RO/EDI/UV) system.